Martini Instruments MI 404 Скачать руководство пользователя страница 1

USER MANUAL

MI 404

Free & Total Chlorine

Meter

Dear Customer,
Thank you for choosing a Martini product. This
manual will provide you with the necessary infor-
mation for the correct use of the instrument.
Please read it carefully before using the meter.

This instrument is in compliance with 

 directives.

Range

0.00 to 5.00 mg/L Cl

2

Resolution

0.01 mg/L (0.00-3.50 mg/L);
0.10 mg/L (above 3.50 mg/L)

Precision

±0.04 mg/L @ 1.50 mg/L

Light Source

Tungsten lamp

Light Detector

Silicon Photocell and 525 nm
narrow band interference filter

Method

Adaptation of the USEPA
method 330.5 and Standard
Method 4500-Cl G.

Environment

0 to 50°C (32 to 122°F);
100% RH max.

Battery Type

1 x 9 volt

Auto-Shut off

After 10' of non-use

Dimensions

192 x 104 x 52 mm
(7.5 x 4.1 x 2”)

Weight

380 g

SPECIFICATIONS:

MEASUREMENT PROCEDURE:

1• Turn the meter on by pressing ON/OFF.

2• When the LCD displays “- - -”,

the  meter is ready.

3• Fill the cuvet with 10 mL of

unreacted sample, up to the
mark, and replace the cap.

4• Place the cuvet into the holder

and ensure that the notch on
the cap is positioned securely
into the groove.

5• Press ZERO and “SIP” will blink

on the display.

6• After a few seconds the dis-

play will show “-0.0-”. The
meter is now zeroed and
ready for measurement.

 9• Immediately reinsert the cuvet into the

holder and ensure that the notch on the
cap is positioned securely into the groove.

10• Press READ and “SIP” will blink during

measurement.

11• The instrument directly displays concentra-

tion in mg/L of free chlorine.

12• Then,  for total chlorine measurement, re-

move the cuvet and add to the reacted
sample 3 drops of DPD 3. Replace the cap
and swirl gently.

13• Reinsert the cuvet into the holder and en-

sure that the notch on the cap is posi-
tioned securely into the groove.

14• Wait for 2 minutes and

30 seconds then press
READ and “SIP” will
blink during measure-
ment.

15• The instrument directly displays concentra-

tion in mg/L of total chlorine.

Interferences

Bromine (positive error)

Chlorine dioxide (positive error)

Iodine   (positive error)

Oxidized Manganese (positive error)

Chromium (positive error)

Ozone (positive error)

 7• Remove the cuvet.

 8• For free chlorine measurement: add 3

drops of DPD 1 indicator and 3 drops of
DPD 2 buffer to an empty cuvet.

Immediately add 10 mL of unreacted sample.
Replace the cap and swirl gently.

ISTMI404  03/05