Soil pH151 Instructions

 

Before first use

 

 

Precipitation  of  the  sensor  electrolyte  may  cause  white 

crystals to form around the cap. This is normal and does not 

affect the performance of the sensor.

 

 

Conditioning

 

 

Condition  the  sensor  before  use.  Remove  the  transparent 

cap  and  fill  with  water  (tap  water)  to  a  height  of  20mm 

(marked  on  the  cap  supplied).  Place  the  cap  on  the  flat 

surface and insert the sensor to soak for one hour. This will 

activate the sensor and dissolve any crystals present. Rinse 

the cap with water afterwards.

 

 

Calibrating the pH Tester

 

 

 

Select  pH  7.01  calibration  solution  for  general  testing, 

select  pH  4.01  for  acidic  and  pH  10.01  for  alkali 

solutions.

 

 

 

Press the ON/OFF button to power on.

 

 

 

Immerse  sensor  in  20mm  of  buffer  solution  and  stir 

gently.

 

 

 

Wait for displayed value to stabilise.

 

 

 

Press the CAL button to enter the calibration sequence.

 

 

 

CAL  will  flash  momentarily  and  then  show  a  flashing 

default reading.

 

 

 

To  complete  calibration,  press  HOLD/ENT  button  to 

confirm.  To  abort  calibration,  press  CAL  button  to 

escape.

 

pH Testing

 

 

 

Take two level 2ml scoops of soil and place in a sample 

container. Add 10ml of deionised water.

 

 

 

Replace cap on container and shake gently for 1 minute.

 

 

 

Transfer the 10ml sample into the transparent cap of the 

pH tester.

 

 