Error Messages 
 

Low Battery 
The scale’s alkaline AA type batteries are flat; 
please replace the batteries. 

 
 
 
 

Overload 
This indicates that the scale’s load sensor(s) 
have been overloaded. Reduce the loading 
and retry. 

 

Counting Error 

1.

 

The signal from the load cells is too 
high. Please remove any weight from 
the scale and try to power on again. If 
the scale continues to show the error 
message, it indicates a fault with the 
electronics or wiring. 

2.

 

The signal from the load cells is too 
low. Please remove any weight from 
the scale and try again. If the scale 
continues to show the error message, 
it indicates a fault with the electronics 
or wiring. 

 

High/Low Zero Count 

1.

 

The scale is above its zero range. 
Please remove any weight from the 
scale and power on again. If the scale 
continues to show the error message, 
it indicates a fault with the 
electronics. 

2.

 

The scale is below its zero range. 
Check there is nothing jammed 
underneath the scale and power on 
again. If the scale continues to show 
the error message, it indicates a fault 
with the electronics.   

 
 

 

EEPROM Error 
This indicates there is a fault with the scale’s 
software and is normally caused by a fault 
with the load cell or wiring. Contact your 
local service representative.   

 

 

Page 18 

Содержание M-560

Страница 1: ...USER MANUAL M 560 Please take time to read these instructions before starting to use the scale Version 1 0 01 17...

Страница 2: ...er Supply and Low Battery 6 Operation Basic Functions 7 Operation Advanced Functions 9 Using the Scale with a Printer 13 Using the Scale with Bluetooth 14 EMC Guidance and Manufacturer s Declaration 1...

Страница 3: ...be exceeded Product Specification Model M 560 Accuracy Class n a Capacity 250kg Weight of Scale Approximately 6kg Units of Measure Kg lb st lb Function Keys ON OFF HOLD TARE BMI UNIT PRINT 0 9 Stabil...

Страница 4: ...ards Batteries should be kept away from small children If swallowed promptly seek medical assistance If you have any problems contact Marsden your local dealer or your service partner Cleaning We reco...

Страница 5: ...n of manufacturer of medical product including address Electro medical appliance with attachment of type B Dispose of old appliances separately from your household waste This product must be disposed...

Страница 6: ...V 2A into the port on the side of the indicator Installing Replacing the Battery Pack 1 Take out the battery housing 2 The rechargeable battery pack will slide into or out of the housing 3 Check that...

Страница 7: ...the Scale Press the ON OFF button firmly The scale will first test all of the display segments The scale will now show its current software version number The scale will now go into weighing mode and...

Страница 8: ...le to Zero If for any reason the scale shows a reading other than 0 0 it can be reset to zero Press the ZERO button once The scale will return to 0 0 Switching between g kg lb and lb st readings You c...

Страница 9: ...once Allow the patient to stand on the scale After a few seconds the scale will lock on the person s weight When the patient steps off the scale the weight will remain on the display Press HOLD again...

Страница 10: ...ess ZERO to confirm the height NB There will always be an active flashing digit in the height display unless HOLD is pressed Weigh the patient as normal The display will show the weight height and BMI...

Страница 11: ...reme left digit will flash Enter the pre set tare value by using the numeric buttons then press TARE again to confirm the value Press ZERO to return to normal weighing mode To use the Tare function ad...

Страница 12: ...period to be edited is selected by pressing the HOLD key E g To input 25 December 2008 8 00 a m Enter the year Press HOLD to confirm and access the date editing field Enter the date E g 12 25 for Dece...

Страница 13: ...I result can be printed Once the person has been weighed and their BMI calculated simply press PRINT to produce the following ticket Connecting the TP 2100 Thermal Printer Plug the cable to the printe...

Страница 14: ...sing the HOLD button select ON enable or OFF disable Press TARE to confirm the setting Note Disabling the Bluetooth function when not in use will reduce battery power consumption Display the bluEt men...

Страница 15: ...that it is used in such an environment Immunity Test IEC 60601 Test Level Compliance Level Electromagnetic Environment Guidance Electrostatic discharge ESD IEC 61000 4 2 6 kV contact 8 kV air 6 kV co...

Страница 16: ...Electromagnetic environment guidance Conducted RF IEC 61000 4 6 Radiated RF IEC 61000 4 3 3 Vrms 150 KHx to 80 MHz 3 V m 80 MHz to 2 5 GHz 3 Vrms 3 V m Portable and mobile RF communications equipment...

Страница 17: ...ons equipment and the M 560 This scale is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled The customer or the user of the scale can help prevent ele...

Страница 18: ...he scale and try again If the scale continues to show the error message it indicates a fault with the electronics or wiring High Low Zero Count 1 The scale is above its zero range Please remove any we...

Страница 19: ...Manufacturer s Declaration of Conformity Page 19...

Страница 20: ...09 364293 E mail sales marsdengroup co uk Manufacturing and Distribution Unit 7 Centurion Business Park Coggin Mill Way Rotherham S60 1FB Head Office Unit 1 Genesis Business Park Sheffield Road Rother...

Отзывы: