Error Messages
Low Battery
The scale’s AA type batteries are flat; please
replace the batteries.
Overload
This indicates that the scale’s load sensor(s)
have been overloaded. Reduce the loading
and retry.
Counting Error
1.
The signal from the load cells is too
high or too low. Please remove any
weight from the scale, switch the
scale off and power on again. If the
scale continues to show this message,
it indicates a fault with the electronics
or wiring.
High/Low Zero Count
1.
The scale is above its zero range.
Please remove any weight from the
scale and power on again. If the scale
continues to show the error message,
it indicates a fault with the
electronics.
2.
The scale is below its zero range.
Check there is nothing jammed
underneath the scale and power on
again. If the scale continues to show
the error message, it indicates a fault
with the electronics.
AD Error
This indicates there is a fault with the scale’s
software and is normally caused by a fault
with the PC board. Contact your local
service representative.
Negative weighing
The weight reading is below -2kg. Press
ZERO to return to 0.0.
Page 16