EMC Guidance and Manufacturer’s Declaration
Guidance and manufacturer’s declaration - electromagnetic emissions.
The M-545 is intended for use in the electromagnetic environment specified below. The
customer or user of this scale should assure that it is used in such an environment.
Emission Test
Compliance
Electromagnetic environment-
guidance
RF emissions
CISPR 11
Group 1
This scale uses RF energy only for
its internal function. Therefore, its
RF emissions are very low and not
likely to cause any interference in
nearby electronic equipment.
RF emissions
CISPR 11
Class B
This scale is suitable for use in all
establishments, including domestic
establishments and those directly
connected to the public low-
voltage power supply network that
supplies buildings used for
domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/flicker
emissions
IEC 61000-3-3
Compliance
Guidance and manufacturer’s declaration – electromagnetic immunity.
The M-545 is intended for use in the electromagnetic environment specified below. The customer or the user
of this scale should ensure that it is used in such an environment.
Immunity test
IEC 60601 test
level
Compliance level
Electromagnetic environment-
guidance
Electrostatic discharge(ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%
Electrical fast transient/burst
IEC 61000-4-4
± 2kV for power
supply lines + 1kV
for input/output
lines
± 2kV for power
supply lines Not
applicable
Mains power quality should be that
of a typical commercial or hospital
environment.
Surge IEC 61000-4-5
± 1kV line(s) to
line(s) ± 2kV line(s)
to earth
± 1kV differential
mode Not
applicable
Mains power quality should be that
of a typical commercial or hospital
environment.
Page 13
Содержание M-125
Страница 17: ...Notes Page 17...