EMC Guidance and Manufacturer’s Declaration
Guidance and manufacturer’s declaration-electromagnetic emissions
The MEDICAL SCALE M-110 is intended for use in the electromagnetic environment specified
below.
The customer or the user of the MEDICAL SCALE M-110 should assure that it is used in such an
environment.
Emission test
Compliance
Electromagnetic environment-guidance
RF emissions CISPR 11
Group 1
The MEDICAL SCALE M-110 uses RF
energy only for its internal function.
Therefore, its RF emissions are very low
and are not likely to cause any
interference in nearby electronic
equipment.
RF emissions CISPR 11
Class B
The MEDICAL SCALE M-110 is suitable
for use in all establishments, including
domestic establishments and those
directly connected to the public low-
voltage power supply network that
supplies buildings used for domestic
purposes.
Harmonic emissions IEC 61000-
3-2
Class A
Voltage fluctuations /flicker
emissions IEC 61000-3-3
Compliance
Guidance and manufacturer’s declaration-electromagnetic immunity
The MEDICAL SCALE M-110 is intended for use in the electromagnetic environment specified
below.
The customer or the user of the MEDICAL SCALE M-110 should assure that it is used in such an
environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment-
guidance
Electrostatic
discharge(ESD)
IEC 61000-4-2
± 8 kV contact
± 2 kV, ± 4 kV, ± 8
kV, ± 15 kV air
± 6 kV contact
± 8 kV air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the relative
humidity should be at least
30%
Electrical fast
transient/burst IEC
61000-4-4
± 2kV for power
supply lines + 1kV
for input/output
lines
± 2kV for power
supply lines Not
applicable
Mains power quality should
be that of a typical
commercial or hospital
environment.
Surge IEC 61000-4-
5
± 1kV line(s) to
line(s) ± 2kV line(s)
to earth
± 1kV differential
mode Not applicable
Mains power quality should
be that of a typical
commercial or hospital
environment.
Page 17