Validation & Calibration
Page
42
of
110
ATTENTION
Cleaning of the G6110 calibration chamber, PAH chamber bushing and the optic(s)
It is crucial that the calibration chamber, PAH chamber bushing for G6110 and the optic(s)
is/are clean prior to validation. Any contaminants present in the chamber, PAH chamber
bushing or on the lens will result in a faulty validation/calibration.
It is vital to wipe the sensor dry prior to and between any validation & calibration attempt.
VALIDATION/CALIBRATION LIQUID SELECTION
Module Type
Calibration Liquid µg/l
Validation
Liquid µg/l
Low Range
Calibration
High Range Calibration
Low
Range
High
Range
Low
Calibration
High
Calibration
Low
Calibration
Middle
Calibration
High
Calibration
G6110
Low
Range
0
100
-
-
-
50
-
G6111
G6110
High
Range
-
-
0
400
800
-
400
G6111
G6110
Dual
Range
0
100
100
400
800
50
200
G6111
-
Table 6-1: Validation/Calibration Liquid Selection
Validation Tolerance µg/l
Standard Concentration
Lower Tolerance
Upper Tolerance
50
45
55
100
95
105
200
160
240
400
360
440
800
760
840
Table 6-2: Validation Tolerance Table
ATTENTION
Do not leave the sensor in the PAH calibration liquid more than 15 seconds since UV light
degrades the PAH calibration liquid.