9.9 Ventilatory settings
Setting range
Factory set default values
(Standard configuration)
Parameter
Adult
Pediatric
Adult
Pediatric
7 - 12
3 – 15
7
3
Maximum apnea time in
Automode (s)
1 - 15
0.5 - 15
4
1
Breath cycle time, SIMV (s)
4 - 100
4 - 150
15
30
Respiratory rate (b/min)
ON/OFF
ON/OFF
ON
ON
Circuit compensation
Note: Circuit
compensation is not
available in NIV.
0 - 2
0 - 2
1.6
1.6
Flow trigger level in
invasive modes, (l/min)
Note: Flow trigger is not
available in NIV.
1:10 - 4:1
1:10 - 4:1
1:2
1:2
I:E ratio
1:10 - 4:1
1:10 - 4:1
1:2
1:2
I:E ratio in backup
1 - 70
1 - 70
30
30
End inspiration (% of peak
flow)
10 - 70
10 - 70
30
30
End inspiration (% of peak
flow) in NIV
0 - 20
0 - 20
5
5
Inspiratory rise time (%)
0 - 0.4
0 - 0.2
0.15
0.15
Inspiratory rise time (s)
0 - 0.4
0 - 0.2
0.4
0.2
Inspiratory rise time (s) in
NIV
—
—
100
80
Maximum permitted
absolute pressure (cmH
2
O)
—
—
62
62
Maximum permitted
absolute pressure in NIV
(cmH
2
O)
0.5-60
0.3 - 20
6
2.4
Minute volume (l/min)
ON/OFF
ON/OFF
OFF
OFF
Nebulizer
144
SERVO-air Ventilator System v2.2, User's Manual
| 9 |
Technical data
|
Содержание SERVO-air
Страница 1: ...x User s Manual SERVO air Ventilator System v2 2...
Страница 2: ......
Страница 4: ...4 SERVO air Ventilator System v2 2 User s Manual Infologic 1 48 TABLE OF CONTENTS...
Страница 16: ...16 SERVO air Ventilator System v2 2 User s Manual 1 Introduction...
Страница 84: ...84 SERVO air Ventilator System v2 2 User s Manual 5 Ventilation modes...
Страница 110: ...110 SERVO air Ventilator System v2 2 User s Manual 6 Ventilatory settings and functions...
Страница 132: ...132 SERVO air Ventilator System v2 2 User s Manual 8 Service Settings...
Страница 168: ...168 SERVO air Ventilator System v2 2 User s Manual 10 Index...
Страница 169: ...x...
Страница 171: ...SERVO air Ventilator System v2 2 User s Manual...