7.4.4 Technical error alarms
Note: Most technical problems require the
attention of service personnel.
High priority alarms
Alarm management checklist
Possible causes
Error code number
Replace the ventilator immediately.
Internal power failure.
1 - 6, 75 - 82
Replace the ventilator immediately.
Control system error.
10, 11
Replace the ventilator immediately.
Expiratory flowmeter error.
37, 40001-40011
Replace the ventilator immediately.
Internal error.
7, 60
Replace the ventilator immediately.
Internal memory error.
42, 44, 10003
Replace the ventilator immediately.
Internal communication error.
16, 25, 35, 43, 55,
20005
Replace the ventilator immediately.
Timeout error.
8, 9
Replace the ventilator immediately.
Barometric error
38, 39
Replace the ventilator immediately.
Monitored value not within range.
40
Replace the ventilator immediately.
Very high turbine temperature
62
Replace the ventilator immediately.
Inspiratory valve error
7
Replace the ventilator immediately.
Inspiratory flow meter error
74
Replace the ventilator immediately.
Turbine communication error
63, 83
Replace the ventilator immediately.
Turbine failure
64 - 69
122
SERVO-air Ventilator System v2.2, User's Manual
| 7 |
Alarm handling
|
Содержание SERVO-air
Страница 1: ...x User s Manual SERVO air Ventilator System v2 2...
Страница 2: ......
Страница 4: ...4 SERVO air Ventilator System v2 2 User s Manual Infologic 1 48 TABLE OF CONTENTS...
Страница 16: ...16 SERVO air Ventilator System v2 2 User s Manual 1 Introduction...
Страница 84: ...84 SERVO air Ventilator System v2 2 User s Manual 5 Ventilation modes...
Страница 110: ...110 SERVO air Ventilator System v2 2 User s Manual 6 Ventilatory settings and functions...
Страница 132: ...132 SERVO air Ventilator System v2 2 User s Manual 8 Service Settings...
Страница 168: ...168 SERVO air Ventilator System v2 2 User s Manual 10 Index...
Страница 169: ...x...
Страница 171: ...SERVO air Ventilator System v2 2 User s Manual...