ROTAFLOW with ICU Kit
|
5 Safety
|
43
|
ROTAFLOW with ICU Kit | User’s Manual | English | 1.0
Copyright MAQUET Cardiopulmonary AG, Hirrlingen
13
Set the required flow.
If the error message
[SIG!]
is still displayed afterwards, follow the safety
procedures and use the ROTAFLOW Emergency Drive (
5.4
DISCONNECTION FROM A HEART-LUNG MACHINE DURING
OPERATION
You can proceed as follows in an emergency:
1
If level monitoring is selected for the ROTAFLOW System, override it on the
heart-lung machine.
2
Connect the level sensor to the “CLS” socket on the ROTAFLOW Console.
3
Switch off the pump switch on the console of the heart-lung machine.
The ROTAFLOW System automatically switches to battery operation.
4
Select “STAND AL” mode for the ROTAFLOW-System (
4.2 “Selecting settings”, page 33).
The ROTAFLOW System may stop.
5
Press the “Bubble override” button to reset the bubble sensor (briefly, not so as
to override the function).
6
If no level sensor is connected, press the “Level override” button for three
seconds. The button flashes.
The ROTAFLOW System starts again.
7
Connect the power cord to the power socket of the ROTAFLOW Console.
8
Switch on the mains power switch at the rear of the ROTAFLOW Console.
9
Remove the ROTAFLOW Console.
10
Release the securing mechanism of the protective plug cap on the underside of
the ROTAFLOW Console and close the protective plug cap.
Disconnection
Steps to be performed
Monitoring system
1 – 6
Monitoring system and power supply
1 – 8
Monitoring system, power supply and
physical
1 – 11