30
|
3 Connection and Installation
|
ROTAFLOW with ICU Kit
|
3.8
REMOVAL
3.8.1
REMOVAL WITH OPERATION AS A STAND-ALONE DEVICE
1
Switch off the on/off switch of the ROTAFLOW Console.
2
Open the lock on the ROTAFLOW Drive and remove all the ultrasonic contact
cream with a lint-free cloth.
3.8.2
REMOVAL FROM A HEART-LUNG MACHINE
1
Switch off the on/off switch of the ROTAFLOW Console.
2
Switch off the pump switch on the console of the heart-lung machine.
3
Remove the ROTAFLOW Console.
4
Release the securing mechanism of the protective plug cap on the underside of
the ROTAFLOW Console and close the protective plug cap.
5
Fit the protective cap to the connection socket of the heart lung machine.
6
Open the lock on the ROTAFLOW Drive and remove all the ultrasonic contact
cream with a lint-free cloth.
6 “Cleaning and disinfection after each use”, page 45.
WARNING!
If you are not using the ROTAFLOW Console, switch off using the on/off switch.
If you only remove the power cord, switch off the mains power switch at the rear
of the ROTAFLOW Console or the pump switch of the heart-lung machine, you
will only be disconnecting the external power supply. In such cases, the
ROTAFLOW Console will default to battery operation. The ROTAFLOW Console
remains switched on, so discharging the batteries.
ATTENTION!
Fit covers to any connection sockets not required on the console of the heart-
lung machine to protect them from liquids.