JBA Instruction Sheet       Issue 1.6  05.03.07         www.mantracourt.co.uk    [email protected]    tel +44 (0)1395 232020                 

 

 

 

JBA 

Load Cell Active Junction Box 

Instruction Sheet  

Features                                          

 
 
 

 

The Strain Gauge Junction Box is used to sum the outputs from up to 
4 Strain Gauges to allow them to be connected to the LCA15/ADW15 
Strain Gauge Indicator/Controller 

 

The individual channel gains can be set up via DIL switches and 
preset potentiometers to allow for 2, 3, or 4 Strain Gauges 

 

Gain is not interactive and offset is preset, to speed up matching of 
Strain Gauge gain 

 

         

 

Introduction

 

The load cell Junction Box is used to sum the outputs up to 4 load cells to allow them to be connected to an ADW15. 
 
The individual channel gains can be set up via DIL switches and preset potentiometers to give an overall gain of unity when 1, 
2, 3 or 4 load cells are connected (e.g. when 2 load cells are used each channel as a gain of 0.5) 
 
The switch setting diagram inside the JBA assumes that the load cell channels are filled starting from No.1 through to No.4 as 
required. Unused channels should be linked out (+IN to - IN). 

Switch settings 

Number of 
Strain 
Gauges 
Connected 




SW1-1 
 
 
 

ON 
OFF 
ON 
OFF 

SW1-2 
 
 
 

ON 
ON 
OFF 
OFF 

SW1-3 
 
 
 

OFF 
OFF 
ON 
OFF 

SW1-4 
 
 
 

OFF 
ON 
OFF 
OFF 

SW1-5 
 
 
 

OFF 
OFF 
ON 
OFF 

SW1-6 
 
 
 

OFF 
OFF 
OFF 
OFF 

SW1-7 
 
 
 

OFF 
OFF 
OFF 
OFF 

SW1-8 
 
 
 

OFF 
OFF 
OFF 
OFF 

Gain Range 
(via preset) 
 
 

x 1 – 0.5 
x 0.33 – x 0.5 
x 0.25 – 0.33 
x 0.20 – 0.25 

 
The unit is designed for 4 wire Strain Gauges, should 6 wire Strain Gauges be 
used, their excitation and sense wires should be both connected to the 
appropriate  'E' terminals. 
The 4 channels can be matched by adjusting the 'Channel Gain' potentiometers 
having first set the DIL switches for the number of Strain Gauges used. 
If access to individual Strain Gauges is possible eg before the platform or hopper 
is in position, then calibration can be carried out by placing a weight on one of 
the cells, and noting the change in display reading on the ADW15.  

 
Repeat this for each remaining Strain Gauge, and adjust the ‘Channel Gain' 
potentiometers, to give the same change in display reading for each cell used. 
 
Should the platform already be in position it will be necessary to use a millivolt 
source to carry out the calibration. Apply a voltage of 10 times the millivolt/volt 
figure given for the appropriate Strain Gauge, to each channel in turn, adjusting 
the 'Channel Gains' to give equal changes in display readings for each cell used. 

Specifications 

Input 
 
Channel Gain  
 
Dimensions 

Up to 20mV/V 
 
x 1 to x 0.25 to allow for 1-4 load cells, DIL switch selectable with potentiometer for fine trim. 
 
200 x 120 x 75mm ABS

 