Peristaltic Pump Series SR

®

 

 
 

SR25.2, SR25.2-G, SR25.2-W 
SR25.3, SR25.3-G, SR25.3-W 
SR25.6, SR25.6-G, SR25.6-W 

 

Instruction Manual 
Version 1.01.01

 

 
 
 
 
 

 

 

 
 
 
 

 

Содержание SR Series

Страница 1: ...Peristaltic Pump Series SR SR25 2 SR25 2 G SR25 2 W SR25 3 SR25 3 G SR25 3 W SR25 6 SR25 6 G SR25 6 W Instruction Manual Version 1 01 01...

Страница 2: ...don t hesitate to contact M C or your M C authorized distributor You will find all the addresses in the appendix of this manual For additional information about our products and our company please go...

Страница 3: ...Maintenance 17 15 1 Changing the pump tubing 18 15 2 Changing contact pulleys and springs 19 15 2 1 Reassembly of the driver 20 15 3 Cleaning the pump head 21 16 Repair information 21 17 Spare parts l...

Страница 4: ...closely Read this manual carefully before commissioning and operating the device If you have any questions regarding the product or the application please don t hesitate to contact M C or your M C au...

Страница 5: ...ening the device for access Make sure that all extern power supplies are disconnected Operate the device only in the permitted temperature and pressure ranges For details please refer to the technical...

Страница 6: ...ttention indicates that an unintended result or situation can occur if the corresponding information is not taken into account Note Note indicates important information relating to the product or high...

Страница 7: ...emperatures requires wearing protective gloves Wear safety glasses Protect your eyes while working with chemicals or sharpe objects Wear safety glasses to avoid getting something in your eyes Wear pro...

Страница 8: ...Terminal 1 5 mm2 Terminal 1 5 mm2 1 x M20 x 1 5 Terminal 1 5 mm2 Dimensions HxWxD 84 x 64 x 52 mm plus 40 mm for the motor 3 31 x 2 50 x 2 04 plus 1 56 for the motor 126 x 105 x 108 mm 4 98 x 4 13 x 4...

Страница 9: ...nt occupational safety regulations during operation Warning Disconnect power supply before opening the device for access Make sure that all external power supplies are disconnected Aggressive condensa...

Страница 10: ...tor DN 4 6 also allows connecting PTFE tubing The peristaltic pump SR25 X X can be equipped with a power supply of 230 V 50 Hz or 115 V 60 Hz Danger Type SR25 X has open electrical wiring Risk of deat...

Страница 11: ...com SR25 X X 1 01 01 11 Figure 2 Drilling pattern for SR25 X Metric dimensions mm are rounded Inch dimensions are for reference only In case of doubt or conflict metric units take priority Figure 3 Di...

Страница 12: ...is a complete pre installed unit Please remove the peristaltic pump carefully from the packaging Check the scope of the delivery specified on the delivery note Please make sure that you have received...

Страница 13: ...ntilated to prevent any accumulation of heat For outdoor installation the pump must be installed in a housing protected from frost in the winter and sufficiently ventilated in summer Exposure to direc...

Страница 14: ...connections for sample gas inlet and outlet the connections are marked accordingly Check for tightness of all sample lines after connecting Connect the tubing to the regular tubing connectors as foll...

Страница 15: ...lasses Wear proper protective clothing 12 2ELECTRICAL CONNECTIONS Connect the cables for the power supply as follows Figure 6 Supply voltage connection for SR25 X W and SR25 X G The supply voltage of...

Страница 16: ...over current protection EN 60335 1 A device to separate the motor from the power supply needs to be provided as part of the electrical installation EN 60335 1 The pump must be installed in such a way...

Страница 17: ...that all external power supplies are disconnected Aggressive condensate possible Media residues in tubing Chemical burns caused by aggressive media possible Wear protective gloves and protective glass...

Страница 18: ...connectors Press the two contact pulleys and check whether the spring pressure is still sufficient if not the contact springs have to be changed see chapter 15 2 Put the new tubing set with the tube...

Страница 19: ...the shaft bore Figure 9 Disassembly of pump head and driver Remove the pump head from the motor shaft Now the driver can be removed from the pump head and is ready for maintenance The removal of the...

Страница 20: ...gs and will replace without any problems the initial springs The replacement springs guarantee full functionality when all four springs are replaced Make sure that contact pulleys move easily on the a...

Страница 21: ...synthetic rubber parts Use oil free compressed air to clean the parts if available Aggressive condensate possible Media residues in tubing Chemical burns caused by aggressive media possible Wear prot...

Страница 22: ...C 1 2 4 90 P 1006 Tubing set with PVDF tube nipples 6mm C 1 2 4 90 P 1020 Roll carier driver SR25 complete S 1 1 90 P 1010 1 set 4 pcs contact springs SR25 for driver R 1 2 2 90 P 1015 1 set 4 pcs rei...

Страница 23: ...www mc techgroup com SR25 X X 1 01 01 23 Figure 11 Spare parts Drawing No 2435 1 07 0...