Ref.: WB-VL-SM Ver.1 3.doc
26
4.5.
DISTANCE ROD TEST
1.
Once Distance measuring test is selected, move the Distance measuring rod
from Home position to maximum length and ensure the count increments from 0
to 90 (±1 count) in steps of 1 in the window. Now return the rod to home position
& ensure the count is displayed as 0:
Fig. 49
2.
Press
key to return to
OPTIONS
.
4.6.
SELF TEST
1.
Once Self test is selected, press
twice to execute the Self test. Status of
respective functions will be displayed as shown below:
Fig. 50
2.
Once all functions are found to OK, press
key to return to
OPTIONS
.
4.7.
SENSOR TEST
1.
Once Sensor test is selected, apply a mild load by pressing the empty Rotor. The
variation in the Millivolt of SA & SB will be displayed as shown below:
Fig. 51
2.
Press
key to return to
OPTIONS
.