Vacuum Degasser User manual
Page 20
MAN 0376
Specifications
General:
Dimensions:
Height: 5.0 in.
Width: 2.87 in.
Depth: 9.81 in.
Weight
6 lb
Channels
3 independent
Degassing Process
Gas permeation through a fluoropolymer membrane
Maximum Flow Rate
10 mL/min.
Degassing Capacity
~2 ppm at 1 mL/min.
Dead Volume
~480 microliters per channel for standard channel
Materials contacting
titrant/solvents
PEEK, Glass-filled PTFE, Teflon AF®, PTFE
Power:
Power requirement using
supplied AC Adapter
100 to 240 VAC, 1A, 50 to 60 Hz
Interchangeable adaptor
plugs
4 supplied with AC Adapter
Interchangeable to AC Adaptor: North America/
Japan, U.K., Continental Europe, Australia
Installation Over-Voltage
Category
II (IEC 60664)
Validation Output:
Signal
5 mVDC / 1 mm Hg absolute from 20 to 800 mm Hg
(0.100 VDC at 20 mm Hg; 4.000 VDC at 800 mm Hg)
Accuracy
±1.0% of reading ±0.010 VDC from 20 to 800 mm Hg
Operating
Conditions:
Ambient Temperature
10 to 35 ºC
Ambient Relative
Humidity (RH)
20 to 80 % RH (without condensation)
Altitude
0 to 2000 Meters
Indoor vs. Outdoor Use
Indoor
Pollution Degree
2 (IEC 60664
MAN0376-2.0 - Vacuum Degasser.book Page 20 Friday, January 12, 2007 2:49 PM