Vacuum Degasser User manual
Page 6
MAN 0376
Identifying the features of the Degasser
Front Panel Connections
ill 7567
There are 3 degassing channels labelled A to C on the front of the Degasser. Each
channel consists of a pair of female connectors which provide the input and output
ports for running the titrant lines through the Vacuum Degasser.
Note
Flow direction is not critical.
Plugs are provided to seal the ports of any unused channels. This must be done to
ensure the Degasser works correctly.
Front Panel Indicators
Three LEDs are located on the front of the instrument above the titrant inlets and
outlets; these indicate the following.
Power (Green)
Indicates when power is applied to the Vacuum Degasser - cable is plugged in and
Power switch turned ON.
Status (Yellow)
Indicates when vacuum level is outside the acceptable operating range. This LED
will light at initial power-up and remain on during pump-down. It will turn off
after a few minutes when the vacuum level goes below 100 mm of Hg absolute.
If an error condition occurs, this LED will flash in one of two modes:
POWER
STATUS
VACUUM
LEDs
Titrant line
I/P / O/P ports
A
B
C
Front of
Vacuum
Degasser
i
MAN0376-2.0 - Vacuum Degasser.book Page 6 Friday, January 12, 2007 2:49 PM