IMPORTANT: This guide is provided as a convenience and for General Information only. Never
use this product without clearly and thoroughly understanding the most recent revision of the
INOmax DS, INOmax DS
IR
and INOvent Operation Manuals. The Operation Manual is the source
for specific, updated information regarding warnings, cautions, checklists, diagrams, and/or
instructions contained in this guide.
See set up diagram on the reverse side of this document
For assistance contact Technical Support 877-566-9466
WARNING:
1. Avoid recirculation of gases.
Undesired recirculation of gases
will occur if fresh gas flows are less
than the patient minute volume and
may result in:
• Higher NO
2
levels due to the
limited ability of the carbon
dioxide absorbent to remove
NO
2
.
• Higher concentrations
than those set due to NO
recirculated through the
absorber.
• Reduction in O
2
concentrations
because nitrogen is the
balance gas for nitric oxide
and will be present in the re-
circulated gases.
2. If the Injector Module was used
in the wet/humidified part of the
breathing circuit, it should be
sterilized between each patient use.
Caution:
1. Note airflow direction on the
Injector Module: the flow out of the
absorber must pass through the
Injector Module in the direction of
the arrow on the module.
2. Nitrous oxide (N
2
O) will affect
the Set NO versus the measured
NO value. For a 50% N
2
O, 50%
O
2
composition, the measured
NO value will be approximately
7% indicated less than the
same Set NO value of 100%
O
2
. For example, at a Set NO
of 20 ppm, measured NO will be
approximately 18 ppm.
3. Similarly, the effect of 2%
v/v isoflurane will result in a
high measured NO value of
approximately 3% indicated for the
same Set NO value at 100% O
2
.
4. Sudden changes in anesthetic
agent concentration may cause
brief transient changes in the
measured NO and NO
2
values.
Connection to a Circle Anesthesia Ventilator System