2. Safety Symbols

 

This symbol adjacent to another symbol, terminal or operating device 
indicates that the operator must refer to an explanation in the Operating 
Instructions to avoid personal injury or damage to the meter.

 

!

This symbol adjacent to one or more terminals identifies them as being
associated with ranges that may, in normal use, be subjected to 
particularly hazardous voltages. For maximum safety, the meter and its 
test leads should not be handled when these terminals are energized.

Equipment is protected by double or reinforced insulation

MAX

Indicates the terminal(s) so marked must not be connected to a circuit 
where the voltage with respect to earth ground exceeds the maximum 
safety rating of the meter.

2.1. Safety Category Ratings

Category
Rating

Brief
Description

Typical Applications

CAT II

Single phase 
receptacles 
and connected
loads 

  Household appliances, power tools

  Outlets more than 10m (30ft) from a CAT III 

  source

  Outlets more than 20m (60ft) from a CAT IV 

  source

CAT III

Three phase 
circuits and 
single phase 
lighting circuits 
in commercial 
buildings 

•  Equipment in fixed installations such as 3- phase 
  motors, switchgear and distribution  panels 
•  Lighting circuits in commercial buildings
•  Feeder lines in industrial plants
•  Any device or branch circuit that is close to a 
  CAT III source

The measurement category (CAT) rating and voltage rating is determined by a 
combination of the meter, test probes and any accessories connected to the meter 
and test probes. The combination rating is the LOWEST of any individual 
component.

2.2. Warning

Operation is limited to CAT II applications when the insulated tips are removed 
from one or both test probes. Refer to Input Limits section in this manual for 
maximum voltage ratings. 

CAT IV 600V
CAT III 1000V

TEST LEADS

5

Содержание MT25

Страница 1: ...25 YEAR SPECIAL EDITION MT25 600V AC DC MULTIMETER INSTRUCTION MANUAL ...

Страница 2: ...2 ...

Страница 3: ...X MIN Button 7 5 4 Flashlight Button 7 5 5 HOLD Backlight Button 7 5 6 AC DC Voltage Measurement 8 5 7 Frequency Duty Cycle Measurements 8 5 8 AC Current Measurement 8 5 9 DC Current Measurements 9 5 10 Resistance Measurements 9 5 11 Diode Test 9 5 12 Continuity 10 5 13 Capacitance Measurements 10 5 14 Temperature Measurements 10 5 15 Battery Replacement 11 5 16 Fuse Replacement 11 6 Maintenance 1...

Страница 4: ...uit under test Use only Major Tech approved test leads with the proper safety category rating Comply with all applicable safety codes Use approved personal protective equipment when working near live electrical circuits particularly with regard to arc flash potential Use caution on live circuits Voltages above 30V AC 42V AC peak or 60V DC pose a shock hazard Do not use if the meter or test leads ap...

Страница 5: ...cal Applications CAT II Single phase receptacles and connected loads Household appliances power tools Outlets more than 10m 30ft from a CAT III source Outlets more than 20m 60ft from a CAT IV source CAT III Three phase circuits and single phase lighting circuits in commercial buildings Equipment in fixed installations such as 3 phase motors switchgear and distribution panels Lighting circuits in co...

Страница 6: ...ymbols Used on LCD Display Degress Farenheit Degrees Celcius 6 micro 10 3 milli 10 3 kilo 10 6 mega 10 Overload Auto Power Off Low battery Autoranging Display hold Maximum Minimum Volts Amperes Alternating current Direct current Minus sign Hertz frequency Percent duty cycle Ohms Continuity Diode test Farads capacitance 9 nano 10 V A Hz F N Ω F C m k M OL AUTO HOLD MAX MIN µ 1 10 2 3 4 5 7 8 9 AUTO...

Страница 7: ...eading and will update when a higher max occurs 2 Momentarily press the MAX MIN button to view the lowest reading The MIN indicator meter will appear on the LCD display The meter will display and hold the minimum reading and will update when a lower min occurs 3 Press and hold the MAX MIN button to end MAX MIN and return to normal operation NOTE The meter does not auto range when the MAX MIN mode ...

Страница 8: ... lead into the input jack COM and the red test lead into the input jack V 4 Touch the test lead probes to the circuit under test 5 Read the frequency or duty cycle on the LCD display 5 8 AC Current Measurements 1 Set the rotary function switch to the 10A position 2 Insert the black test lead into the COM input jack and the red test lead into the 10A input jack 3 Remove power from the circuit under...

Страница 9: ...ears on the LCD display 3 Insert the black test lead into the COM input jack and the red test lead into the Ω input jack 4 Touch the test lead probes to the component under test If the component is installed in a circuit it is best to disconnect one side before testing to eliminate interference with other devices 5 Read the resistance in on the LCD Display Never test resistance on a live circuit W...

Страница 10: ...red test lead into the input jack Ω 4 Touch the test lead probes to the capacitor under test 5 Read the capacitance value on the LCD display It may take up to a minute to get a stable reading on large capacitors 1 Set the rotary function switch to the position F C 2 Press the button to select readings in or MODE F C 3 Connect the Temperature Probe to the Banana Plug Adapter Note the and markings o...

Страница 11: ...endable service if the following care instructions are performed 5 16 Fuse Replacement To avoid electric shock remove the test leads from the meter before removing the battery fuse cover WARNING To avoid electric shock do not operate meter until the battery fuse cover is securely fastened to the meter WARNING 6 Maintenance 1 KEEP THE METER DRY If it gets wet wipe it off 2 USE AND STORE THE METER I...

Страница 12: ...rcuit voltage 2 8V DC typical Audible signal if the resistance is approx 50Ω or less Minus symbol is displayed for negative polarity 0 C to 40 C 32 F to 104 F 80 relative humidity 20 C to 60 C 4 F to 140 F 80 relative humidity 147 x 68 x 50mm 318g Complies with IEC 61010 1 v 3 for measurement Category lll 600V Pollution Degree 2 Class 2 Double insulation is displayed 4000 count LCD display OL is d...

Страница 13: ...n 600V DC or 600V AC RMS Resolution 1mV 10mV 600V 1V Range 4 000A 10 00A Accuracy of reading digits 3 0 5 digits Function AC Current Overload Protection 10A 600V Fuse AC Response 50 to 60Hz Resolution 1mA 10mA Range 4 000A 10 00A Accuracy of reading digits 3 0 5 digits Function DC Current Overload Protection 10A 600V Fuse Resolution 1mA 10mA 999 9Hz 1 2 3 digits 0 1Hz 9 999kHz 1Hz Range 9 999Hz 99...

Страница 14: ...its Function Capacitance Input Protection 600V AC RMS or 600V DC Pulse Width 100µs 100ms Frequency Range 5Hz to 150kHz Sensitivity 8V RMS Resolution 10pF 100pF 400 0µF 0 1µF 4000mF 1µF Input Protection 250V AC RMS or 250V DC 7 Range 20 to 760 C 4 to 1400 F Accuracy of reading digits 3 5 C Function Temperature Resolution 1 C 1 F 3 9 F 14 Range 0 1 to 99 9 Accuracy of reading digits 1 2 2 digits Fun...

Страница 15: ...15 ...

Страница 16: ...www major tech com www majortech com au sales major tech com info majortech com au MAJOR TECH PTY LTD South Africa Australia ...

Отзывы: