4.4. DC Current Measurements
Caution:

 Do not make current measurements on the 20A scale for 

longer than 30 seconds. Exceeding 30 seconds may cause 
damage to the meter and/or the test leads.

1.

Insert the black test lead banana plug into the negative (COM) 
jack.

2.

For current measurements up to 200mA DC, set the function 
switch to the mA range and insert the red test lead banana plug 
into the (mA) jack.

3.

For current measurements up to 20A DC, set the function switch 
to the A position and insert the red test lead banana plug into the 
20A jack.

4.

Remove power from the circuit under test, then open up the 
circuit at the point where you wish to measure current.

5.

Touch the black test probe tip and the red test probe tip in series 
to the circuit whose current you want to measure.

6.

Apply power to the circuit.

7.

Read the current on the display. The display will indicate the 
proper decimal point value.

4.5. AC Current Measurement

Warning:

 To avoid electric shock, do not measure AC current on any 

circuit whose voltage exceeds 250V AC.

Caution:

   Do not make current measurements on the 20A scale for 

longer than 30 seconds. Exceeding 30 seconds may cause 
damage to the meter and/or the test leads.

1. 

Insert the black test lead banana plug into the negative (COM) 
jack.

2. 

For current measurements up to 200mA AC, set the function 
switch to the mA range and insert the red test lead banana plug 
into the (mA) jack.

3. 

For current measurements up to 20A AC, set the function switch 
to the A position and insert the red test lead banana plug into the 
20A jack.

4. 

Remove power from the circuit under test, then open up the 
circuit at the point where you wish to measure current.

5. 

Touch the black test probe tip and the red test probe tip in series 
to the circuit whose current you want to measure.

10

Содержание MT1883

Страница 1: ...MT1883 Manual Ranging Multimeter INSTRUCTION MANUAL...

Страница 2: ...urement 9 4 3 AC Voltage Measurement 9 4 4 DC Current Measurement 10 4 5 AC Current Measurement 10 4 6 Resistance Measurement 11 4 7 Continuity Check 11 4 8 Diode Test 11 4 9 Frequency Measurement 12...

Страница 3: ...t under test before cutting unsoldering or breaking the circuit Small amounts of current can be dangerous Use caution when working above 60V DC or 30C AC RMS such voltages pose a shock hazard When usi...

Страница 4: ...on which if not avoided may result in damage to the product This symbol advises the user that the terminal s so marked must not be connected to a circuit point at which the voltage with respect to ear...

Страница 5: ...lay with function indication Polarity Automatic negative polarity indication Overrange OL mark indication Low Battery Indication The is displayed when the battery voltage drops below the operating lev...

Страница 6: ...50V and 20A 250V fuse Maximum Input 200mA DC or 200mA AC RMS on mA ranges 20A DC or AC RMS on 20A range Range Resolution Accuracy 200 0mV 0 1mV 2 000V 1mV 20 00V 10mV 200 0V 100mV 1000V 1V 0 5 of rdg...

Страница 7: ...250V AC RMS Range Resolution Accuracy 2 000mA 1uA 200 0mA 100uA 20 00A 10mA 2 0 of rdg 3 dgts 1 2 of rdg 3 dgts 3 0 of rdg 10 dgts Range Resolution Accuracy 200 0 0 1 2 000k 1 20 00k 10 200 0k 100 2 0...

Страница 8: ...RMS 4 Operation Warning Risk of electrocution High voltage circuits both AC and DC are very dangerous and should be measured with great care If 1 appears in the display during a measurement the value...

Страница 9: ...and value If the polarity is reverse the display will show minus before the value 4 3 AC Voltage Measurements Warning Risk of electrocution The probe tips may not be long enough to contact the live p...

Страница 10: ...r to the circuit 7 Read the current on the display The display will indicate the proper decimal point value 4 5 AC Current Measurement Warning To avoid electric shock do not measure AC current on any...

Страница 11: ...the display The display will indicate the proper decimal point value 4 7 Continuity Check Warning To avoid electric shock never measure continuity on circuits or wires that have voltage on them 1 Set...

Страница 12: ...y The digital reading will indicate the proper decimal point value 4 10 Capacitance Measurements Warning To avoid electric shock disconnect power to the unit under test and discharge all capacitors be...

Страница 13: ...omes at or drops below the operating voltage BAT will appear in the right hand side of the LCD display The battery should be replaced 2 Follow instructions for installing the battery See the battery i...

Страница 14: ...y item under test 2 Open the back cover by loosening the screw on the back cover using a phillips head screwdriver 3 Remove the old fuse from its holder by gently pulling it out 4 Install the new fuse...

Страница 15: ......

Страница 16: ...www major tech com www majortech com au sales major tech com info majortech com au MAJOR TECH PTY LTD South Africa Australia...

Отзывы: