11
The NCV function works on any rotary switch position.
1. Test the detector on a known live circuit before use.
2. Hold the top of the meter very close to the voltage source as shown.
3. If voltage is present, the back light will flash a bright red.
4.1. NON-CONTACT VOLTAGE (NCV)
Low battery indication
a battery " "symbol is displayed when the
battery voltage drops below the operating
level.
Measurement rate
3 times per second nominal.
Auto power off
Meter automatically shuts down after approx.
15 minutes of inactivity. To disable the auto
power off feature, hold down the MODE Button
and turn the meter on.
Operating environment 0°C to 50°C (32°F to 122°F) at <70% relative
humidity.
Storage temperature
-20°C to 60°C (-4°F to 140°F) at <80%
relative humidity.
Relative humidity
90% (0°C to 30°C); 75% (30°C to 40°C); 45%
(40°C to 50°C);
For inside use,
Operating: 3000m, Storage
max height:
10,000m.
Pollution degree
2
Safety
The instrument complies with IEC/EN
61010-1:2001 and IEC/EN 61010-031 :2002
Power
One 9V battery, NEDA 1604, IEC 6F22.
Dimensions
182 (H) x 82 (W) x 55 (D) mm
Weight
Approx.: 375g.
Accuracy
Accuracy is given at 18°C to 28°C
(65°F to 83°F), less than 70% RH
Function
Range