Section 5 Operation Instructions
1. Place Genius Dry Bath Incubator on a sturdy and level surface in a safe, dry
place, away from laboratory traffic.
2. Ensure that the AC power switch is OFF, then plug the three-pronged power
cord into a grounded three-prong AC outlet of the appropriate voltage (115V
or 220V as indicated on the rating sticker near the AC cord on the back of
the unit).
3. Select suitable module block(s) or appropriate water volume and put it /
them into the Genius Dry Bath Incubator.
4. Turn the AC power ON.
5. Run temperature calibration procedure when using the instrument for the
first time (see page 12).
6. Press
or
Key to adjust to the desired temperature.
7. If setting heating time is required, press
Key, and then press
Key or
Key to adjust timer upon your request. The unit will stop with
alarm when timer is up.
8. Press the
Key to start heating.
9. If to reset timer is required during heating, press
Key to deactivate
heating.
10. If the optional external temperature probe is used, plug it into the connector
located on the rear panel.
Note:
If you’d like to have an additional external temperature probe,
please contact with Major Science and send the instrument back to
our factory.