![Maico MB 11 Classic Скачать руководство пользователя страница 45](http://html1.mh-extra.com/html/maico/mb-11-classic/mb-11-classic_operation-manual_3710975045.webp)
Operation Manual MB 11 Classic
8100498-4_OM_MB11_Classic_3.18_EN_16a.docx
43
01/16
6.4.5 Test result "Refer"
If 100% of the pass criteria is not reached after 180 seconds of data
collection time, the test is automatically terminated and a “REFER” will be
displayed below and to the right of the graph.
In this case, follow your screening program protocol to determine whether
another screening should be performed or whether the baby will be referred
for a diagnostic follow-up test.
When you leave the test screen by clicking on the button “QUIT”, the test
result will be stored automatically in the data base.
Figure 21
MB 11–program Test result “REFER”