2.4 Electrical Safety and Measuring Security
This icon indicates that patient applied parts of the instrument
conform to IEC 60601-1 Type B requirements.
The protection class of the system depends on the used
power supply.
In Case of Emergency
In case of emergency, disconnect the instrument from the
power supply.
Do not position the instrument in a way that it is difficult to
operate the disconnection device. The supply mains and the
power socket shall be accessible at all times.
Do not use the instrument if the mains cable and/or the outlet
is damaged.
Safety against electrical hazard is guaranteed only when the
connected notebook computer is powered by batteries
respectively the computer´s power supply accords to the
IEC 60601-1 or IEC 60950-1 safety regulations.
External devices such as computer, printer or Ethernet which
are connected to the device must meet electrical safety
requirements, such as IEC 60601-1 series for medical
electrical equipment or IEC 60950 for IT equipment. This is
to avoid electrical shock to the user or patient.
This complete system consisting of the MAICO device,
computer and isolation transformer, is suitable for use in the
patient environment. Equipment not complying with
IEC 60601 shall be kept outside the patient environment, as
defined in the standard (at least 1.5 m from the patient).
It is possible to use IEC 60601-1 certified USB and Ethernet
insulators which allows the electrically safe use of electrical
medical devices in medical electrical systems.
Any person who connects external equipment has created a
system and is therefore responsible for the system complying
with the requirements of IEC 60601-1-1. If in doubt, contact
your service technician or local representative for help.
The instrument is not intended for operation in areas with an
explosion hazard.
Never short-circuit the terminals.
WARNING
WARNING
WARNING