IntroductIon

The MVD 100 measuring Amplifier is designed to control forces, loads, pressures, etc. in connection 
with analog sensors, predominantly with DMA transmitters.  After conditioning the actual value 
signal, the device is digitally assisted by a processor (MSP 430/TI).  A total of 3 switching thresholds, 
which  include  hysteresis,  switching  performance,  and  pick-up  delay,  can  be  programmed  and 
generated.

2. 

descrIptIon

Low drift precision operational amplifiers are used to amplify the sensor voltage to a level which can 
be processed by the processor’s own A/D converter.  The digitalized actual value is then compared 
with several set values.  Switching signals are triggered if the actual value coincides with, or exceeds 
any of these set values.  The relevant set values are entered via a keyboard with LCD display which 
can be plugged into the amplifier’s PCB.  This control unit may remain connected or be removed 
from the amplifier after programming since it does not affect the amplifier’s function. 

Digital measuring Amplifier

Type MVD100

Содержание MVD100

Страница 1: ...n Low drift precision operational amplifiers are used to amplify the sensor voltage to a level which can be processed by the processor s own A D converter The digitalized actual value is then compared...

Страница 2: ...Set the C1 switch to position 2 programming mode The following image will appear in the display S1 xxx This means in normal language The set value of the first threshold S1 amounts to XXX at present...

Страница 3: ...nt value E Adjusted end point value Special feature If 000 is entered for S1 and or S2 and or S3 pressing the K3 transmission key will not result in the 000 value being transmitted to the memory but i...

Страница 4: ...s off at 48 of full load The parameters T1 T2 and T3 constitute the turn on delay times after which the relays are activated 5 Serial interface The serial interface optional of the processor RS 232 al...

Страница 5: ...are determined on the basis of this value x0 becomes 0x push 4 times and then becomes 3x through the addition of 30h 30h is added to 0y in order to obtain 3y Example STX v iii t sss pp ETB 02h 030 11...

Страница 6: ...sis 1 101 Hysteresis 1 100 Delay time 1 101 Delay time 1 100 Delay time 1 10 1 Inversed 1 0 or 1 2nd switching point 102 2nd switching point 101 2nd switching point 100 Hysteresis 2 102 Hysteresis 2 1...

Страница 7: ...1 Plug in card with 32 pin DIN strip 3 Version with 20 pin terminal strip 2 Version with 20 pin terminal strip 8 Plug in card with 32 pin DIN strip incl front panel 2 Switching point 0 Without switchi...

Страница 8: ...8 MVD100 User s manual 7 connection to 230 115 vac supply voltage Figure 7 1 Jump wire connections for voltage supply...

Страница 9: ...4 USA Phone 1 716 668 5555 Fax 1 716 668 8705 E mail magtrol magtrol com Magtrol SA Route de Montena 77 1728 Rossens Fribourg Switzerland Phone 41 0 26 407 3000 Fax 41 0 26 407 3001 E mail magtrol mag...