DES 410 & DES 411

Power Supply

User’s Manual

Via Paolo Uccello 4  -  20148 Milano   

Tel +39 02 48 009 757   Fax +39 02 48 002 070      [email protected]   www.dspmindustria.it

Содержание DES 410

Страница 1: ...DES 410 DES 411 Power Supply User s Manual Via Paolo Uccello 4 20148 Milano Tel 39 02 48 009 757 Fax 39 02 48 002 070 info dspmindustria it www dspmindustria it...

Страница 2: ...ny part of the contents of this manual without the express permission of Magtrol is strictly prohibited Purchase Record Please record all model numbers and serial numbers of your Magtrol equipment alo...

Страница 3: ...onic products are earth grounded to ensure personal safety and proper operation 2 Check line voltage before operating electronic equipment 3 Make sure that dynamometers and motors under test are equip...

Страница 4: ...Revision Date 1st Edition revision A February 2014 Date Edition Changes Section 02 19 14 1st Edition rev A Passing shielded cables into stuffing gland procedure added 2 6 2 02 19 14 1st Edition rev A...

Страница 5: ...UNITS 10 2 6 CONNECTING THE DES 410 and DES 411 Power Supply 11 2 6 1 Passing Unshielded Cables Into Stuffing Gland 11 2 6 2 Passing Shielded Cables Into Stuffing Gland with EMC stuffing gland 12 2 7...

Страница 6: ...the various units in a test bench 10 Figure 2 6 Stuffing gland Overview and separated 11 Figure 2 7 Connect Shiel of EMC Stuffing gland 12 Figure 2 8 Location of fuses F1 F2 F3 SolderLink SL12 Contact...

Страница 7: ...g unit without risk Manual Organization This section gives an overview of the structure of the manual and the information contained within it Some information has been deliberately repeated in differe...

Страница 8: ...s used to draw the operator s attention to information directives procedures etc which if ignored may result in damage to the material being used The associated text describes the necessary precaution...

Страница 9: ...t to the coils of the brake within the dynamometer They are controlled by an electronic peripheral the Magtrol DSP7000 High Speed Programmable Controller Note The DES 41x is intended for use in an ind...

Страница 10: ...he output current to be delivered This excitation current is controlled by a set point in the 0 10VDC range The nominal value of the excitation current is set by internal resistors There are two discr...

Страница 11: ...ce 50 k DIGITAL INPUTS GALVANICALLY INSULATED Remote Control of the Network Input PSC Relay coil 24 VDC 11 mA Control of the Electromagnetic Clutch Optocoupler activated by 24 VDC 2 5 mA Stand by enab...

Страница 12: ...72 190 175 17 219 12 10 90 30 in 11 30 10 71 7 48 6 89 0 63 8 58 0 47 0 39 3 54 1 06 N P Q mm 200 290 15 in 7 87 11 42 0 59 Specifications The DES 410 DES 411 supplies are delivered with integrated ca...

Страница 13: ...rcuit Control Interface Control Inputs Outputs Supply Inputs Outputs Fuse Network Filter Differential Input Selection of Nominal Current Electrical Alarm Excitation Temp Dynamometer Temp Clutch Contro...

Страница 14: ...housing of the power supply are provided in Figure 1 The data sheet provides all other dimensions necessary for the installation of the power supply The housing of the DES 410 and DES 411 has four hol...

Страница 15: ...HE USE OF A 6A 30mA GROUND FAULT CIRCUIT BREAKER IS RECOMMENDED 2 3 MOUNTING WITHOUT COOLING For any DES 410 and for DES 411 Power Supply providing an excitation current up to 5A included dynamometer...

Страница 16: ...m 2 mm It is necessary to mill holes in the plate for the feet of the unit in order to get a good thermal contact using heat sink compound to improve the thermal coupling This plate can then be mounte...

Страница 17: ...10 411 Power Supply Chapter 2 Installation Configuration Figure 2 4 DES 411 Power Supply with cooling plate mounted to the test bench table DES41x Power Supply Metallic case support Cooling plate Test...

Страница 18: ...Supply The test bench is controlled by a DSP 7000 Programmable Controller Figure 2 5 shows the connection between the various units in a test bench DES 410 411 Power Supply DSP7000 High Speed Program...

Страница 19: ...g gland to adapt to various diameters of cable Element Can be removed from element by simply pushing it outwards c Pass the cables through elements if used and d Reassemble the elements of the stuffin...

Страница 20: ...th to the elements output and cut the shield over 5 8 mm as show in Figure 2 7 Pass the stripped cable through the element e Connect the shield on the metallic part of element as show in Figure 2 7 f...

Страница 21: ...of fuses resistors and SolderLink in accordance with the dynamometer model and the Main supply voltage 2 7 1 DES 410 Main Board F1 F2 F3 SL12 SW1 SW2 Location The contacts SW1 and SW2 must be left Op...

Страница 22: ...be left Open to allow the DSP7000 to control the primary supply circuit F3 2 A 250 V 5x20 SW1 SW2 F1 F2 SL12 SL1 SL3 Figure 2 9 Location of fuses F1 F2 F3 SolderLink SL12 Contacts SW1 SW2 on the DES 4...

Страница 23: ...er WB PB DES 41x Type lex A R39 Ohm R102 Ohm 115 V AC F1 F2 230 V AC F1 F2 1WB PB2 7 8 K 410 0 5 348 none T2A T1A 2WB PB2 7 8 K 410 1 1 21K 90 9K T2A T1A 3PB2 7 8 K 410 1 5 2 15K 36 5K T2A T2A 4WB PB2...

Страница 24: ...Dynamometer A temp Earth ground 1 2 3 4 5 6 Shield 7 1 2 3 4 5 6 7 8 9 10 11 12 1 2 3 4 5 Terminal group P4 Terminal group P2 Terminal group P3 Terminal group P1 brown yellow black grey pink red blue...

Страница 25: ...in alarm Main alarm ON 0 V 12 V 1 2 3 4 5 Terminal group P2 N C Alarm switch Current Set Point_C Current Set Point_H Stand by Temperature alarm Coupling Electronic alarm Water Flow N C Primary supply...

Страница 26: ...ion When the DES 410 or 411 Power Supply is purchased as part of a complete motor test system it is calibrated by Magtrol according to the dynamometer with which it will be used Calibration requires s...

Страница 27: ...on Description of the defect and the conditions in which it appeared Description of the test bench drawing photographs sketches etc Description of the tested object drawing photographs sketches etc De...

Страница 28: ...ng Measurement and Control of Torque Speed Power Load Force Weight Tension Displacement Via Paolo Uccello 4 20148 Milano Tel 39 02 48 009 757 Fax 39 02 48 002 070 info dspmindustria it www dspmindustr...