32
Magtrol Model 6400 Torque Transducer Display
Chapter 6 – Calibration
MAINTENANCE
7. Press and release SHIFT. The display appears as follows:
POWER/AUX
TORQUE
SPEED
Figure 6–4 Full Scale Torque Display 1
8. Apply +10.000 VDC. The display appears as follows:
POWER/AUX
TORQUE
SPEED
Figure 6–5 Full Scale Torque Display 2
9. Press and release SHIFT twice to exit calibration and return to main menu.
Note:
To exit calibration mode without making any changes, press the
SHIFT button twice.
6.3.3
A
UXILIARY
I
NPUT
O
FFSET
AND
G
AIN
6.3.3.1
Computer
1. Connect the external voltage reference to the Auxiliary Input BNC connector.
2. Apply 0.000 VDC.
3. Send the CZA command.
4. Apply +10.000 VDC.
5. Send the CFA command.
6.3.3.2
Manual
1. Enable the calibration mode as follows:
1.) Turn instrument power OFF.
2.) Press and hold the LEFT
W
and RIGHT
X
buttons simultaneously while turning the
instrument power ON. The software revision date appears on the display as shown in
Figure 6–1.