21
5.3.7.
Comparator
Up to 8 thresholds can be set for 2 or 4(2 steps / 4 steps) for each set frame.
Outputs the judgment value for each set area.
The set can be specified by the comparator group number setting command.
The following is an example of comparator judgment when setting the number
of steps.
Example of comparator judgment
5.3.8.
Add/Sub function
Addition and subtraction of measured values of two measuring units are possible.
The calculation result is output as measurement data (current value) for each
frame. Up to 16 frames can be set.
The calculation formula is shown below.
Axis calculation result (current value)
=
<Sing1>Measurement value of measuring unit Ⓐ
+
<Sing2>Measurement value of measuring unit Ⓑ
(Example)
Measurement value of measuring unit Ⓐ
= 10[μm]
Measurement value of measuring unit Ⓑ
= 5[μm]
Calculation setting:Sign1 = +, Sign2 = -
Measured value after calculation(Current value) =
(+10) +
(-5)
= 5[μm]