17
5.
Function
5.1.
Functions list
Item
Contents
Condition
Module configuration
Whole system
MG80-PN can be expanded to 255 units via PROFINET network (total 4080axes)
Number of
connectable modules
Counter module MG80-CM
:
From 1 to 16units (Measuring unit 1 to 16 axes)
I/O module LZ80-K1/K2
:
Max. 2units
Resolution
Measuring unit
Input resolution
0.1
μ
m/0.5
μ
m/1
μ
m/2
μ
m/5
μ
m/10
μ
m
Data capture capability
10Mbps
Max. 10000 data/sec.
Data format
7-digit fixed length with sign and decimal point
(
Top zero suppress
)
Peak hold function
Calculate maximum value, minimum value, peak to peak value of each frame
Hold updating peak value during pause period
Start recalculation of peak value by start
Output data type
Single axis
Current value, maximum value, minimum value, peak to peak value of each axis
Addition and
subtraction function
Current value, maximum value, minimum value, peak to peak value of 2-axis add/sub
Comparator
Compare the measurement data of each frame
※
and output the comparator result.
Number of steps
2 steps
/
4 steps
Number of groups
8 groups
Reset
Reset count value to zero
Reference point
When using the reference point of the measuring unit, set the reference point position is
use as the reference value
Preset
Preset to set value
Master preset
Master preset to set value
I/O module
The specified function can be assigned to each terminal of the I/O module. Perform the
operation according to the assigned terminal function.
※
“Frame” is a function that can be set the measured value or calculated
value of the measuring unit in a prepared frame and call it with Explicit
communication. The details are described in 5.2 on the next page.