8
OxyMag_rev26
1.
Description
1.1
Intended Use
Oxymag MAGNAMED’s Transport and Emergency Electronic
Ventilator
–
belong to the family of equipment for ventilatory support
of neonatal, pediatric and adult patients with respiratory failure, with
controlled volume, pressure and time cycled. Intended for use in
patients from neonatal, pediatric, adult and adults with morbid
obesity. Oxymag interacts with the patient through an invasive or
noninvasive interface that supplies ventilator air to the patient's
airway.
Oxymag provides a mixture of ambient air and oxygen at
concentrations adjusted by the operator using the accurate oxygen
concentration System using the venturi principle. O
2
concentration
is obtained through a galvanic cell or optionally from a
paramagnetic cell by indirect contact with the patient's gas by
passing gas through the sensor. In addition, it performs the control
of flows and pressures in the respiratory circuit to provide the
ventilation modalities appropriate to the patient's condition.
The possible ventilation modes of this ventilator are:
•
VCV
–
Volume Controlled Ventilation (can be Assisted);
•
PCV
–
Pressure Controlled Ventilation (can be Assisted);
•
PLV
–
Pressure Limited Ventilation (can be Assisted)
–
Available Weight ≤ 6.0Kg (Neonatal);
•
P-SIMV
–
Synchronized Intermittent Mandatory
Ventilation with Pressure Controlled cycle;
•
V-SIMV
–
Synchronized Intermittent Mandatory
Ventilation with Volume Controlled cycle;
•
CPAP/PSV
–
Continuous Pressure Ventilation with
Pressure Support;
•
DualPAP
–
Ventilation at two CPAP levels (with or
without Support Pressure). Adjustments can be
performed through this modality in order to obtain APRV
mode (Airway Pressure Release Ventilation);
•
Noninvasive ventilation (NIV) by mask can be activated
in all ventilation modes with leakage compensation.
During ventilation in CPAP/PSV a backup ventilation can be established in the case of APNEA; this ventilation can be chosen between
VCV, PCV, PLV or OFF.
•
This device should be operated only by
healthcare professional with expertise
in mechanical ventilation and qualified
and trained in its use.
•
In CPAP/PSV and DUALPAP modes ∆PS
= OFF should be set to deactivate the
support pressure and BACKUP should
be set to OFF to deactivate backup
ventilation. Be aware that when
adjusting parameter BACKUP to OFF
backup ventilation will be INACTIVE
during APNEA.