85
MAGNAMED TECNOLOGIA MÉDICA S/A
11.10.9 APRV
APRV
–
Airway pressure release ventilation
Description:
This mode allows spontaneous cycles on 2
levels of baseline pressure and can be
achieved by appropriate adjustments in
DualPAP mode;
For this mode selects inverted ratio in
DUALPAP. With this adjustment is carried out
a pressure relief airway, obtaining
APRV
–
Airway Pressure Release Ventilation
Set parameters:
•
PR. SUPERIOR
•
T. SUPERIOR
•
PR. INFERIOR
•
T. INFERIOR
•
CONCENTRATION
•
ΔPS (
Pressure Support
– PEEP
)
•
PRESSURE LIMIT;
•
CYCLING FLOW (% FLOW);
•
TRIGGER PRESSURE
•
FLOW PRESSURE
•
RISE TIME
•
FLOW (NEONATAL)
•
BACKUP
o
Backup VCV
▪
VOLUME
▪
RATE
▪
FLOW
▪
LIMIT PRESSURE
o
Backup PCV
▪
PRESSURE INSP
▪
RATE
▪
TIME INSP
▪
RISE TIME
o
Backup PLV-NEONATAL
▪
PRESSURE INSP
▪
RATE
▪
TIME INSP
o
Backup Auto
(1)
1
–
Whenever the apnea set
time is reached, the ventilator
triggers a ventilatory cycle
whose configuration is based
on the current mode settings.