66
MAGNAMED TECNOLOGIA MÉDICA S/A
2. Ventilation with Inspiratory Pause, after the adjusted volume is delivered the ventilator maintains
interrupted expiration until T
INS
is completed. Afterwards, the ventilator cycles for expiration. The
characteristic is the pressure plateau formation (the gap between the peak and the plateau
depends from the airways resistance).
3. If the pressure or flow trigger is activated, the ventilator seeks to synchronize at the beginning of
the next inspiration with the patient’s effort, according to the established levels. The information
related to what type of trigger activated the inspiratory cycle is informed in the status and
messages area. The detection of the patient’s inspiratory effort in order to synchronize occurs at
any moment during the expiratory time.
•
If the patient makes inspiratory efforts and triggers are suitably set, the ventilatory mode becomes
assisted-controlled. In this situation, the monitored breathing frequency can be significantly higher than
the set one.
4. Waveform Flow ASCENDING (or accelerated).
5. Waveform Flow DECENDING (or decelerated).
6. Waveform Flow SINUSOIDAL.
7. Representation of Limited Pressure. In this situation, the ventilator limits the pressure within the
adjusted value and, as a consequence of such factors as the patient’s lung compliance and
imposed pressure limit, the adjusted volume IS NOT DELIVERED, and this condition is informed
on the screen alarm area (message LIMITED PRESSURE).
•
By reaching the pressure limit defined on the Maximum Pressure adjustment (LIMITED
PRESSURE alarm), the Adjusted Volume IS NOT DELIVERED.
•
Default values are only an initial reference. Readjust the ventilation parameters according to the
patient’s needs.
11.10.2 PCV
PCV
–
Pressure Controlled Ventilation