59
MAGNAMED TECNOLOGIA MÉDICA S/A
Average time to recharge to full load
(operation module)
(1)
4,0 hours
15%
(1) The battery should be made at ambient temperature from 5 to 35 ° C
11.4.1.1 External Power Supply DC
Table 13
– External Power source DC
Item
Specification
Tolerance
Power Supply
(1)
Voltage: 12 to 15 V
DC
10%
Current: 4 A
(1) External Power Supply OPCIONAL
•
It is not possible to recharge the internal batteries of the equipment through the external DC
power supply.
•
The sole purpose of this entry is to allow the equipment to be powered temporarily by a
source of compatible external power when there are no other alternatives.
•
In ventilation before disconnecting an external power, make sure that there is enough load
on internal batteries or connect the device to the external power grid.
11.4.2 Connectors
Table 14 - Connectors
Item
Specification
External Power Network (rede)
Connector 3 (tree) plugs, as ABNT NBR 14136:2012
Central pin ground
External Power Supply
Connector housing 3.96mm
–
4 pins
180º female
Distal flow sensor
Connector Redel
–
6 pin female receptacles
Содержание Fleximag