45
MAGNAMED TECNOLOGIA MÉDICA S/A
8
Cleaning, Disinfection and Sterilization
It is important to establish a routine for cleaning, disinfection or sterilization of equipment and its
components.
The following describes the main forms of cleaning, disinfection or sterilization according to the
characteristics of each component and equipment.
Equipment cleaning
8.1.1 External Parts
External ventilator surfaces of Fleximag should be cleaned with a clean, soft cloth moistened with the
enzymatic detergent
•
Be careful to not accumulate residue in the connections of the equipment.
•
For cleaning do not use non-compliant products to polymer.
8.1.2 Components
The components that come in directly contact with respiratory gases must be periodically disassembled
for cleaning, disinfection or sterilization.
8.1.2.1 Respiratory circuit and exhalation valve
Circuits and parts of silicone should be cleaned according the following steps:
8.1.2.1.1
Wash
a) Always use potable water for this procedure;
b) Use neutral and enzymatic detergent. Dilution should be performed as recommended by the
manufacturer.
c) Immerse the entire body of the flow sensor and the silicone line in the detergent solution, keeping
the solution in contact with the accessories for at least 3 minutes;
d) The external parts of the parts should be cleaned with a clean, soft cloth moistened with the
enzymatic detergent. The internal parts must be cleaned by immersion.