40
MAGNAMED TECNOLOGIA MÉDICA S/A
•
Check that the set volume to audio alarms is available with the distance that the equipment
will be from clinical body.
6.5.5 TREND
The equipment records all events occurring in the last 24 hours of ventilation, such as adjusted
ventilatory parameters as well as monitored values and all alarms conditions during this period.
It is possible to see the TREND curves up to 3 ventilation parameters simultaneously from the 13
parameters available for selection.
Once defined the parameters, simply set the interval you want to see the trend.
This interval can range from 15 minutes to 24 hours, with other choices of time within the same period.
The greater the range selected, the longer the time between events, or less details are shown.
Made the settings, just press the ENTER button.
While the machine retrieves the requested data, an hourglass appears and simultaneously the chart
is updated with the current data of selected parameters on the right.
After all the data required to display the trend graph, the hourglass will no longer be displayed and
from this point a cursor that allows the shift in the trend chart appears.