73
•
For there to be enough battery power during a power outage, it is important that the
equipment remains connected to a mains power supply WHENEVER POSSIBLE.
•
For the capacity of the batteries in normal operation to be full, they shall be replaced
as indicated in the technical specification.
•
Replacement of the internal batteries shall be performed by trained and qualified
personnel.
9.5 Water Collectors with Coalescing Filter
•
Do not expose the filter container to non-polycarbonate compatible materials.
•
Replace the filter when it is clogged so that it does not slow down the inlet flow of the
equipment.
To remove accumulated water, simply press the pin found on the bottom of the collector.
To perform filter replacement, consider the following instructions:
Instruction
Image
Remove the screws with an Allen wrench