9.2 Preventive Maintenance Schedule
Magnamed recommends performing preventive maintenance of ICU ventilators with your authorized
network throughout the country.
Below spreadsheet with the schedule of maintenance and exchange of preventive parts. If you need
more details, contact the MagnaService (Magnamed Technical Assistance).
Table 8
– Preventive maintenance schedule
ITEM
Period
5000 h
or 1
year
10000 h
or 2
years
15000 h
or 3
years
20000 h
or 4
years
25000 h
or 5
years
30000 h
or 6
years
35000 h
or 7
years
40000 h
or 8
years
45000 h
or 9
years
50000 h
or 10
years
Assessment according
to procedures
MAGNAMED
x
x
x
x
x
x
x
x
x
Evaluate replacement
of consumables
x
x
x
x
x
x
x
x
x
LI-ION Battery Pack
x
x
x
x
O
2
cell (galvanic)
x
x
x
x
Disposal of equipment
x
9.3 Consumable items
Table 9
– Consumable items replacement
ITEM
Period
Silicon line
2 years or after 50 hot steam sterilization cycles
Diaphragm
2 years or after 50 hot steam sterilization cycles
Adult proximal flow sensor
5 years or after 50 hot steam sterilization cycles
Pediatric proximal flow sensor
5 years or after 50 hot steam sterilization cycles
Neonatal proximal flow sensor 5 years or after 50 hot steam sterilization cycles
Integrated valve
5 years or after 50 hot steam sterilization cycles
Mesh filter
500 h of use
Air/O
2
inlet filter
Replace if it is obstructed
9.4 Internal Batteries
These batteries are responsible for maintaining the operation of the equipment even in the absence of
electricity and its duration in normal operation is specified in chapter 10.