MAMOS – UPGRADE OF ANALYSER’S FIRMWARE 

 

 

Service manual 

Version: 1.2 
12/2022 
 

 

Содержание MAMOS

Страница 1: ...MAMOS UPGRADE OF ANALYSER S FIRMWARE Service manual Version 1 2 12 2022...

Страница 2: ...www madur com User manual maMoS 2 1 PROCEDURE OF FIRMWARE UPGRADE CPU 1 Turn OFF the analyser and disconnect from mains 2 Open the analyser s casing remove 6x blanks and unscrew 6x Pozidriv PZ2 screws...

Страница 3: ...n the CPU board install jumper short circuit the two top pins 4 Turn ON the analyser Analyser will start in programming mode Display will show Waiting for CPU 5 Download WSD program from Analog Device...

Страница 4: ...nfiguration button 7 Select the communication port the one you use to communicate with mamos via PC Connect PC mamos using USB cable 8 Select Erase the CODE ONLY option otherwise all mamos settings wi...

Страница 5: ...rmware mamos3 hex Firmware must match the analyser s hardware configuration loading inappropriate hex file can in the best case scenario disable analyser Please consult every firmware update with madu...

Страница 6: ...www madur com User manual maMoS 6 12 After upgrade program will report the status 13 Turn the analyser OFF 14 Remove the jumper 3 and turn the device ON...

Страница 7: ...F the analyser and disconnect from mains 2 Open the analyser s casing remove 6x blanks and unscrew 6x Pozidriv PZ2 screws 3 To perform Display upgrade special cable is needed Silicon Labs USB Debbuger...

Страница 8: ...www madur com User manual maMoS 8 4 Connect adapter to display PCB to connector called Prog...

Страница 9: ...www madur com User manual maMoS 9 5 Download Silicon Labs Flash Utility Silicon Labs Flash Utility 6 Select USB Debug Adapter 7 Check Power target option 8 Press Connect...

Страница 10: ...www madur com User manual maMoS 10 9 Program confirms it has been connected...

Страница 11: ...www madur com User manual maMoS 11 10 Select tab Download Hex File Go Stop 11 Select hex file usually called Disp4x20 hex 12 Check option Erase all code 13 Press Download...

Страница 12: ...www madur com User manual maMoS 12 14 Firmware is loading...

Страница 13: ...www madur com User manual maMoS 13 15 Program is successfully loaded 16 Close flashing program 17 Disconnect Silicon Labs debugger 18 Put back the analyser 19 Turn the analyser ON...