Manual
maMoS
100/200/300/400
Picture 14.
Analogue outputs on the monitor.
The analogue output (
U1
÷
U4
) signal range is 0 ÷ 5V or 0 ÷ 10V (set with the help
of the
MaMoSII.exe
software), whereas the current output (
I1
÷
I4
) signal range is 0÷20mA
or 4÷20mA (also set with the help of the
MaMoSII.exe
software).
3.3.3. Digital output (PWM)
The monitor has a single digital
PWM
output (
PV3
). This is an open drain output,
the active state of the output means short-circuiting the
PV3
ending
to the ground potential
by internal high-power N-MOS transistor. This can be used for switching an external relay,
etc. The
PWM
output can be controlled via analogue outputs or digital inputs. It can also
proceed to the active state when the monitor has initiated the measurement phase. The
PWM
output function can be defined with the help of the
MaMoSII.exe
software.
15