Device Operation

 

Connecting and Starting the Data Logger

 

1.

 

Once the software is installed and running, plug the 
interface cable into the docking station (IFC400 or 
IFC406). 

2.

 

Connect the USB end of the interface cable into an 
open USB port on the computer. 

3.

 

Place the data logger into the docking station. 

4.

 

The data logger will automatically appear under 

Connected Devices 

within the software. 

5.

 

For most applications, select 

Custom Start 

from 

the menu bar and choose the desired start method, 
reading rate and other parameters appropriate for the 
data logging application and click 

Start

(

Quick Start 

applies the most recent custom start options, 

Batch 

Start 

is used for managing multiple loggers at once, 

Real Time Start 

stores the dataset as it records while 

connected to the logger.) 

6.

 

The status of the device will change to 

Running 

or 

Waiting to Start

, depending upon your start method. 

7.

 

Remove the data logger from the docking station and 
place it in the environment to measure. 

Note: 

The device will stop recording data when the end of memory 

is reached or the device is stopped. At this point the device cannot be 
restarted until it has been re-armed by the computer. 

 

Downloading Data from a Data Logger

 

1.

 

Place the logger into the docking station. 

2.

 

Highlight the data logger in the 

Connected Devices 

list. Click 

Stop 

on the menu bar. 

3.

 

Once the data logger is stopped, with the logger 
highlighted, click 

Download

4.

 

Downloading will offload and save all the recorded 
data to the PC. 

Trigger Settings (Transient Mode)

 

The  PR1000IS  samples  up  to  128  Hz  (7.8  ms)  and  begins 
recording  after  exceeding  user  defined  set  points.  After 
triggering,  the  device  will  record  a  chosen  number  of 
samples (Window Mode) or until reaching the stop set point 
(Two Point Mode). The device can take as many as 380,928 
readings with both channels enabled and 419,020 pressure 
only readings. The data logger will record up to 50 samples 
of “pre-trigger” data. 

1.

 

In the 

Connected 

Devices 

panel, 

click the device 
desired. 

2.

 

On the 

Device 

Tab, 

in the 

Information 

Group, click 

Properties

. Or, 

right-click the 
device and select 

Properties 

in the context menu. 

3.

 

Select 

Trigger 

in the Properties window. 

4.

 

Trigger formats are available in 

Window 

or 

Two Point 

Mode

. Window mode allows a high and/or low trigger set 

point, and a trigger sample count or “window” of time 
recorded when set points are exceeded to be defined. Two 
point allows for different Start and Stop set points to 
be defined for both the high and low triggers. 

Refer to the 

Trigger Settings - MadgeTech 4 Data Logger 

Software 

video o

madgetech.com 

for instructions on 

how to configure Trigger Settings. 

 

 

 

Communication

 

To  ensure  desired  operation  of  the  PR1000IS, 

please  keep  the  surface  clear  of  any  foreign  objects  or 

substances

. The PR1000IS’s data is downloaded through external contact with the IFC400 or IFC406 docking 

station. Covering the surface with foreign objects 

(i.e. Calibration Labels) 

can prevent the communication 

and/or downloading process. 