
"Offset adjust" function
This function is used to “fine tune” a sensor reading whenever a sensor is deployed or reading inaccurately.
For example, a depth sensor may be sitting slightly off the bottom of the stream, or the user has deployed it
above a silt line on the side of a pipe.
1. Click “
Offset adjust
” from the “
Edit channel
” dialogue box and the “
Channel offset adjust
”
dialogue box appears.
The “
Current value
” from the sensor (in this case a depth sensor with frequency output) is displayed
at the top of the window and is updated approximately once per second.
2. To adjust the “
Current value
”, enter your current measured value in the “
Corrected value
” field.
3. Click on the “
Accept
” button to calculate the new “
Offset
” parameter value for the associated channel.
Clicking on the “
Cancel
” button terminates the procedure without altering the offset value.
For a short video of the procedure click "
Play
"-
-
WARNING: When calibrating a "
Depth
" channel, measurements should ALWAYS
be carried out from the bottom of the channel/pipe to the top of the water level
MACE XCi User Manual - 216 -
Содержание HydroMace XCi
Страница 1: ......
Страница 13: ...MACE XCi User Manual 13...
Страница 19: ...MACE XCi User Manual 19...
Страница 33: ...MACE XCi User Manual 33...
Страница 36: ...MACE XCi User Manual 36...
Страница 167: ...NOTE None of these parameters can be edited MACE XCi User Manual 167...
Страница 182: ...MACE XCi User Manual 182...
Страница 187: ...MACE XCi User Manual 187...
Страница 200: ...NOTE In the example the Net flowrate will be the sum of Flow Rate 1 Flow Rate 2 Flow Rate 3 MACE XCi User Manual 200...
Страница 260: ...5 Tighten all three screws 6 Reconnect the battery cable assembly to the backplane board MACE XCi User Manual 260...
Страница 265: ...MACE XCi User Manual 265...