7.5.7.4
Password
This screen is used to configure the Password FTP server.
7.5.7.5
Remote Dir
This screen is used to configure the Remote Directory of
FTP server where files will be uploaded from data-logger.
7.6
Discharge Table Configuration
In order to calculate discharge, we must first configure
the Discharge table in LogPRO TOUCH device. Discharge
table can be easily configured following the steps below:
1.
Create a text file discharge-config.xml and open it
in any simple text editor like Notepad.
<discharge-config>
<levelTag>Level</levelTag>
<surfaceVelocityTag>Velocity
</surfaceVelocityTag>
<map>
<add>0.00,0.00,0.89</add>
<add>0.48,1.76,0.89</add>
<add>0.48,1.76,0.89</add>
<add>0.49,1.82,0.89</add>
<add>0.50,1.87,0.89</add>
<add>0.51,1.92,0.89</add>
<add>0.52,1.97,0.89</add>
<add>0.69,2.87,0.89</add>
<add>0.70,2.93,0.89</add>
<add>0.71,2.98,0.89</add>
<add>0.72,3.03,0.89</add>
<add>0.75,3.19,0.89</add>
<add>0.77,3.30,0.89</add>
<add>0.79,3.43,0.89</add>
<add>0.80,3.49,0.89</add>
<add>0.82,3.62,0.89</add>
<add>0.83,3.69,0.89</add>
<add>0.84,3.76,0.89</add>
<add>0.85,3.84,0.89</add>
<add>0.86,3.92,0.89</add>
<add>0.87,4.00,0.89</add>
<add>0.88,4.07,0.89</add>
<add>0.89,4.15,0.89</add>
</map>
</discharge-config>
Here:
Option
Description
levelTag
Name of Tag that measures
Water Column Level
surfaceVelocityTag Name of Tag that measures
Surface Velocity of Water
map
Cross-section and K factor
mapping for metered water
column.
NOTE: The table has to be in
increasing order of water
column.
<add>0.48,1.76,0.89</add>
•
First value 0.48
represents the water
column height (in meters)
•
Second Value 1.76
represents the cross-
section area at the water
column (in m
2
)
•
Third Value 0.89
represents the K factor at
water column
2.
Prepare to transfer the file in LogPRO TOUCH.
a.
Arrange an empty USB pen drive. Create
following folders in USB Pen drive.
AutoCopy > Flashdisk > Firmware > 1.1.0.0 >
Configs
b.
Inside Config folder, save the discharge-
config.xml and safely eject USB pen drive.
3.
Apply the new discharge table
a.
Insert the USB pen drive in LogPRO TOUCH.
Ensure that device is powered ON.
b.
Wait for 5 seconds and eject the USB pen
drive.
c.
Restart the LogPRO TOUCH.
Only simple FTP supported.
SSL is NOT supported.
Содержание LogPRO TOUCH
Страница 6: ...6 UM200002 ISSUE A REV 00 User Manual 5 Components...
Страница 7: ...7 UM200002 ISSUE A REV 00 User Manual 6 Cable Wiring Diagram LOGPRO TOUCH...
Страница 9: ...9 UM200002 ISSUE A REV 00 User Manual 7 3 Mechanical Figure 3 Physical Dimensions mm...
Страница 16: ...16 UM200002 ISSUE A REV 00 User Manual 8 4 Tripod Dimensions...
Страница 17: ......