Resetting PI counts in regular intervals
Time input is used to reset PI counts in regular time intervals to continuously accumulate and store pulse
counts over certain time period.
Given below are examples of parameter setting for logging PI1 while resetting the counts in a regular interval
by using AI1.
■
Logging pulse counts for 1 minute by resetting PI every minute
Parameter
Setting
AI1
CH setting
Time input
Time input
Second
Alarm zone setting
Partitions
2
Zone 1, Upper limit
10
Zone 2, Lower limit
10
Zone 1, Reset totalized value (PI)
Enable, PI1
PI1
Assign a pulse input channel to be measured.
Logging
Storing rate time unit
Mode
Minute
Min.
1
Pen
Assign a pen with PI1.
■
Logging pulse counts for 1 hour by resetting PI every hour on the hour
Parameter
Setting
AI1
CH setting
Time input
Time input
Minute
Alarm zone setting
Partitions
2
Zone 1, Upper limit
10
Zone 2, Lower limit
10
Zone 1, Reset totalized value (PI)
Enable, PI1
PI1
Assign a pulse input channel to be measured.
Logging
Storing rate time unit
Mode
Hour
Choose hour and
offset (Apply same
setting to all hours (00
to 23)
Logging
Enable
Offset (min.)
0
Offset (sec.)
0
Pen
Assign a pen with PI1.