19
73ET/74ET/75ET USERS MANUAL EM-7353-0003B Rev.5
2.2
USING A BLANK CF CARD FOR 74ET & 75ET
For the 74ET or the 75ET, you need to copy files required to operate the 7xET to your blank CF Card. For the 73ET,
no copying files are required.
Required files are stored under the following directories in the PC Recorder Software Package CD (model:
CHARTLSP2).
74ET: \Software\74ET(English)
75ET: \Software\75ET(English)
Each folder contains an .exe file and a Backup folder. Copy both of them to the CF Card. The property of files in
Backup folder is set to ‘Read Only.’ Remove these settings. (Select ‘Apply changes to this folder, sub-folders and
files.’)
Remark 1: 74ET/75ET Backup folders exist for English and Japanese. Files in 74ET(Japanese) or 75ET(Japanese)
folders are not usable.
Remark 2: With the Windows XP, the ‘Read Only’ setting of the folder is automatically reset when files are copied to
the CF Card. ‘Read Only’ box in the ‘Property’ is marked with
■
, which indicates the folder is NOT ‘Read Only.’
However, files in the folder may still be ‘Read Only.’ In that case, Set ‘Read Only’ box to blank and select ‘Apply
changes to this folder, sub-folders and files.’
2.3
STARTING UP
When the power supply to the 7xET is turned on after the CF Card has been set, Initial Check view appears on the
screen. The view is shown while the 7xET is creating the data file in the CF Card. After that, the initial view as shown
in Figure 2.3 appears on the screen.
Remark: The 7xET does not start up unless the CF Card is inserted before the power is turned on.
Figure 2.3: Initial view while stand-by.
CAUTION !
The 7xET may require a certain time after a control button is touched until the designated action or setting is com-
plete. Operate the 7xET control buttons slowly and confirm each step of operation before going to a next step.
Hasty operation may damage the CF Card.
Содержание 75ET
Страница 24: ...20 7xETBLD USERS MANUAL EM 7353 0003C...