WARNING!!! IMPORTANT
Conditions of Use for Gel Batteries
The battery while installed in the weighing scale must not remain discharged and
must be recharged immediately. Leaving the battery discharged may result in short-
ening its life, reducing its capacity or damaging it.
When not in use, the battery should be recharged regularly, at intervals no longer than
2 weeks.
The battery can be stored and used in temperatures between 0°C and 35°C. Storing
and using the battery outside the indicated temperature range will limit its life or lead
to its
failure.
The waranty covers the replacement of the scale battery free of charge if any factory
defect is found. Factory defects of the battery are considered to be:
- break in internal connections
- internal short circuit
- leakage (not caused by external force)
The warranty is no longer valid if:
- the battery has been stored and used in temperatures below 0°C and above 35°C
- the battery has been kept in a discharged state (voltage below 10.8V) for more than
20 hours
- deep discharge of the battery (voltage below 9.9V)
- the battery has been charged with equipment other than the dedicated charger
- the battery was used to power equipment other than the weighing scale
- mechanical damage to the battery casing is found