Lutron Electronics PDO-519 Скачать руководство пользователя страница 1

Pen type, auto calibration

DISSOLVED 

OXYGEN METER

Model : PDO-519

Your purchase of this
DISSOLVED OXYGEN
METER marks a step
forward for you into the
f i e l d   o f   p r e c i s i o n
measurement.  Although
this METER is a complex
and delicate instrument,
its durable structure will
allow many years of use if
p r o p e r   o p e r a t i n g
techniques are developed.
Please read the following
instructions carefully and
always keep this manual
within easy reach.

       OPERATION MANUAL

Содержание PDO-519

Страница 1: ...d for you into the field of precision measurement Although this METER is a complex and delicate instrument its durable structure will allow many years of use if p r o p e r o p e r a t i n g technique...

Страница 2: ...r the best condition when user receive the DIGITAL OXYGEN METER along the PROBE it should fill the Probe s Electrolyte at first Probe filling Electrolyte Probe head with Diaphragm set The procedures t...

Страница 3: ...robe head with diaphragm set 4 3 6 Battery compartment Cover 4 3 7 Protection cover for probe head 4 4 MEASURING PROCEDURE 5 4 1 Calibration 5 4 2 Dissolved Oxygen DO measurement 6 4 3 Oxygen in Air O...

Страница 4: ...All in one Pen type Dissolved Oxygen Meter easy operation Automatic Temp compensation from 0 to 50 for sensor Microprocessor circuit assures high accuracy and provides special functions and features M...

Страница 5: ...reference only Temperature 0 8 1 5 Sensor The polarographic type oxygen probe with Structure an incorporated temperature sensor Temp 0 to 50 Compensation Automatic adjustment Memory Records Maximum Mi...

Страница 6: ...ssories Soft carrying case CA 05A 1 PC Included Operation manual 1 PC Spare Probe head with Diaphragm OXHD 04 2 set Probe filling Electrolyte OXEL 03 1 set Optional Spare Probe head with probe and Dia...

Страница 7: ...T PANEL DESCRIPTION Fig 1 3 1 Display 3 2 Power button 3 3 HOLD button Unit button 3 4 REC button button 3 5 Probe head with diaphragm set 3 6 Battery compartment Cover 3 7 Protection cover for probe...

Страница 8: ...utton 3 2 Fig 1 The Display 3 1 Fig 1 will show the O2 and the Temp value as example 21 4 26 4 O2 The upper display O2 display value will be changed wait for approx 3 minutes at least until the upper...

Страница 9: ...tion procedures under wide and ventilating environment for best effect 4 2 Dissolved Oxygen DO measurement 1 After the meter be calibrated above procedure 4 1 the meter is ready for measurement 2 Pres...

Страница 10: ...id is recommended In this way errors due to the diffusion of the oxygen present in the air in the solution are reduced to a minimum 5 Rinsed the probe accurately with normal tap water after each serie...

Страница 11: ...on 3 4 Fig 1 once to start the Data Record function and there will be a REC symbol on the display With the REC symbol on the display a Press the REC button 3 4 Fig 1 once the REC MAX symbol along with...

Страница 12: ...is run out or the diaphragm probe head with diaphragm set exist problem dirty If yes please fill the electrolyte or change the Probe head with diaphragm set and make the new calibration The considerat...

Страница 13: ...ust be replaced in the following way Probe filling Electrolyte OXEL 03 Probe head with Diaphragm set 1 Unscrew the Probe head 5 3 Fig 2 2 Pour out the old Electrolyte from the container of the Probe h...

Страница 14: ...nstrument become inaccurate 2 To replace the battery rotate and remove the Battery Cover 3 1 Fig 1 take out the old batteries install new batteries DC 1 5V battery UM 4 AAA x 4 PCs 3 When install the...

Отзывы: