Lutron Electronics ESF-106 Скачать руководство пользователя страница 1

ELECTROSTATIC FIELD

METER

Model : ESF-106

Your purchase of this 
ELECTROSTATIC FIELD
METER  marks a step
forward for you into the
field of precision
measurement. Although
this METER is a complex
and delicate instrument,
its durable structure will
allow many years of use
if proper operating
t e c h n i q u e s   a r e
developed. Please read
t h e   f o l l o w i n g
instructions carefully
and always keep this
manual within easy
reach.

       OPERATION MANUAL

Содержание ESF-106

Страница 1: ...the field of precision measurement Although this METER is a complex and delicate instrument its durable structure will allow many years of use if proper operating t e c h n i q u e s a r e developed...

Страница 2: ...3 10 SENSE disc 3 3 11 Magnetic SPACER 3 3 12 RS 232 output terminal 3 3 13 RESET button 3 3 14 GROUND terminal 3 3 15 Stand 3 3 16 Battery Cover Compartment 3 3 17 EARTH wire 3 3 18 Alligator clip 3...

Страница 3: ...equivalent 2 SPECIFICATIONS 2 1 General Specifications Display LCD size 51 mm x 30 mm Max indication 19999 LCD with backlight On Off Circuit Custom one chip of microprocessor LSI circuit Measurement...

Страница 4: ...al 1 PC included 1 Magnetic Spacer 1 PC Ground wire 1 PC Ground CLIP 1 PC Carrying case CA 06 1 PC Optional USB cable USB 01 accessories RS232 cable UPCB 02 Data Acquisition software SW U801 WIN Excel...

Страница 5: ...ton 3 4 HOLD button 3 5 RECORD button MAX MIN 3 6 ZERO button 3 7 button 3 8 ALARM button 3 9 button 3 10 SENSE disc 3 11 Magnetic SPACER 3 12 RS 232 output terminal 3 13 RESET button 3 14 GROUND term...

Страница 6: ...rical outlet ground terminal 2 Operator is attached to the EARTH ground by way of an Anti static wrist strap not provided and finger placed on the GROUND point 3 2 Fig 1 4 3 Zero procedure Before you...

Страница 7: ...n 3 9 Fig 1 to decrease the value 3 When you finish the first figures you can press the ALARM button 3 8 Fig 1 to setting next figures 4 When you finish your limit value you can press the ALARM button...

Страница 8: ...econds The display will revert to the current reading 4 8 LCD Backlight 1 You can turn ON the Backlight for to help see the display When use the instrument in a dark place or under the SUN 2 While the...

Страница 9: ...sconnect from any circuit and the instrument is power OFF 7 RS232 PC SERIAL INTERFACE The instrument has RS232 PC serial interface via a 3 5 mm terminal 3 12 Fig 1 The data output is a 16 digit stream...

Страница 10: ...1 LSD D8 MSD For example If the display reading is 1234 then D8 to D1 is 00001234 D0 End Word 0D RS232 FORMAT 9600 N 8 1 Baud rate 9600 Parity No parity Data bit no 8 Data bits Stop bit 1 Stop bit 8 B...