84
ALM LEV: see the content of this chapter parameter area the SPO2 set " ALM LEV ".
ALM REC: see this chapter parameter area the SPO2 set ―ALM REC‖.
SWEEP: see this chapter waveform area the SPO2 set "SWEEP‖.
WAVE COLOR:.see this chapter waveform area the SPO2 set "WAVE COLOR‖
BRAIN ALARM(intelligent alarm): 10, 25, 50, 100 seconds, disable, for example, intelligent alarm
range is set to 50, then when NELLCOR oximeter alarm limit 97, and the lower limit is 90, and the
measured oxygen value is 99, then from exceeding the alarm limit commences consecutive exceeds
the alarm limit of 25 seconds as soon as they sound and light alarm at the same time the oxygen
values next to the circle and draw back to square one. The intelligent alarm to reduce false alarms,
allow doctors to more accurate and timely master oxygen changes. (This feature is only effective
NELLCOR oxygen)
SPO2 ALM SETUP: see the content of this chapter parameter area the SPO2 set alarm settings.
PR ALM SETUP: see the content of this chapter parameter area the SPO2 set alarm settings.
If the oxygen probe in the machine NELLCOR probe "SPO2 alarm settings, PR (pulse rate) alarm settings
set upper and lower limits of the alarm limit.
DEFAULT: Select this option to enter the the SPO2 default configuration dialog box, the user can
choose "yes" or "no" to "will be the default configuration" or "original configuration will be set.
13.8
Alarms & Reminders
SpO
2
alarm information
When the alarm record function under certain menus is on, those physical alarms caused because relevant
parameters exceed the specified alarm limits will automatically output alarm parameter values and relevant
measurement waveforms.
Physical alarms:
Reminders
Causes
Alarming levels
SPO2 over-high
SpO2 measurement value higher than the set
upper limit
User defined
SpO2 over-low
SpO2 measurement value lower than the set
lower limit
User defined
PR over-high
PR measurement value higher than the set
upper limit
User defined
PR over-low
PR measurement value lower than the set
lower limit
User defined
Technical alarms
Reminders
Causes
Alar
m
Level
s
Remedial Measures