TEMP Monitoring
15.1
TEMP Monitoring
The monitor has two TEMP measuring channels. The temperature data can be measured with the TEMP
detector. Teperature is not a standard configuration for the neonatal monitor.
15.1.1
TEMP measurement setting
For one-time TEMP detectors, users must insert the TEMP cables into slots and then connect the detectors
with such cables; for reusable TEMP detectors, users can directly insert them into slots.
Closely paste TEMP detectors with patient body.
Turn on the system power supply.
Warning
Before monitoring users should check status of detector cables by plugging out the TEMP detector
cable from the hole, then the screen will display the error information “T sensor disconnected” and
make voice alarming.
Attention
A one-time TEM detector can only be used once.
Warning
Be careful to use or store TEMP detector and cables; spare detectors and cables should be
wrapped into loose rolls. Tough wires inside the detector and cables, if any, may cause
mechanical injury.
Warning
Calibration of a TEMP detector must be done for every two years or comply with your
hospital‟s specified schedule. When requiring calibration, please contact the manufacturer.
Attention
During monitoring process a TEMP detector will self-detect once per hour; such self-detection
last for 2 seconds and will not affect normal work of the TESP monitor.
15.2
TEMP setting in parameter area
Users can use the knob to move the cursor onto the TEMP hotkey in the parameter area and press the knob
to enter the TEMP Setting menu.