Troubleshooting
37
Sample Probe
Problems with the sample probe can lead to fluid leaks and
pressurization problems, as well as inhibit sample acquisition.
Symptom
Possible problem
Solution
Sample probe leaks.
The sample probe is clogged.
Clean the sample probe. See
page 20.
Sample arm is stuck in the up
position.
The system isn’t properly
pressurized.
Check pressure settings. Ensure
that the sample probe is not
clogged and there are no leaks
in the syringe seal or syringe
valve.
Sample arm is stuck in the down
position.
The sample probe height is too
low, or the path to the well is
blocked.
DO NOT turn off the Luminex
XYP instrument.
1. Remove the blue light
housing from the analyzer.
2. Unscrew the Cheminert
fitting. The system monitor
changes from “Busy” to
“Running” and sample
acquisition continues. Click
Cancel
to make adjustments
before continuing with the
rest of the samples.
3. If the sample arm does not
raise, save the data that has
been collected. Turn off the
analyzer, but keep the XYP
instrument on.
4. Exit the Luminex software.
5. Turn the analyzer back on
and restart the software.
6. Run a Wash with distilled
water to remove air from
the system.
7. Verify probe height.
Содержание Luminex 200
Страница 1: ...Luminex 200 System User Manual Lum nex...
Страница 9: ...iii Luminex 200 System User Manual Hardware 47 Software 48 xMAP Reagents 49...
Страница 10: ......
Страница 42: ......
Страница 56: ...Luminex 200 System User Manual 46...
Страница 60: ...Luminex 200 System User Manual 50...
Страница 64: ......