Luminex 200 System User Manual
36
Fluid Leaks
Fluid leaks can result in poor pressurization and failed sample
acquisition.
Regulator not adjusted properly.
If using bottles, open the center
door on the Luminex analyzer.
Use a screwdriver to adjust the
regulator to fit in the center of
the green region on the Run
Batch tab.
Symptom
Possible problem
Solution
Pressure too low.
The sample probe is clogged.
Clean the sample probe. See
page 20.
The syringe seal leaks.
Replace the syringe seal. See
page 24.
The syringe valve leaks.
Hand-tighten the syringe
connection (silver knob) on the
syringe valve. Run a Prime
command. If leaks continue,
contact Technical Support.
Large amount of fluid pooled
around instrument.
Fittings or fluid lines are
damaged.
Contact Technical Support.
Fluid dripping from the sample
probe.
The sample probe is clogged.
Clean the sample probe. See
page 20.
The sample three-way valve is
faulty.
Contact Technical Support.
Fluid is leaking from the front
of the analyzer.
The syringe seal leaks.
Replace the syringe seal. See
page 24.
The syringe valve leaks.
Hand-tighten the syringe
connection (silver knob) on the
syringe valve. Run a Prime. If
leaks continue, contact
Technical Support.
Содержание Luminex 200
Страница 1: ...Luminex 200 System User Manual Lum nex...
Страница 9: ...iii Luminex 200 System User Manual Hardware 47 Software 48 xMAP Reagents 49...
Страница 10: ......
Страница 42: ......
Страница 56: ...Luminex 200 System User Manual 46...
Страница 60: ...Luminex 200 System User Manual 50...
Страница 64: ......